2.2.3 Remote installation
Signal converter in
19" insert
6
2. Installation of signal converter
1. Fit the SENSORPROM
SENSORPROM
®
2. Mount the guide rails in the rack system as shown. Distance between guide rails is 21 TE.
Guide rails are supplied with the rack system and not with the signal converter.
3. Mount the connection board as shown. The left side of the connection board must flush to the
left of the guide rail.
4. Connect the cables as shown under "Electrical connection".
5. Plug the signal converter into the rack system.
SFIDK.PS.028.M1.02
®
unit on the connection board supplied with the signal converter. The
unit is supplied with the sensor.