Part Number: 701627
Calibration:
Advanced Display
(High resolution))
Welch Allyn
Service Documentation
Follow the operating instructions for the BB-3200 or 9600 PLUS closely.
The measured value must be within:
NOTE:
The measured temperature will vary with ambient temperature.
Calibration mode must be activated: the display is flashing and showing
- - -
"CAL" and "
" alternately.
One temperature measuring with probe cover must be done.
Push once the
- High resolution mode: the display is flashing and showing "hr" and
measured temperature e.g "736" (37.36 = hundredth ° C) alternately.
The first digit will be not showing.
Push once again the
- Ambient temperature mode: the display is flashing and showing "AT"
and ambient temperature e.g. "23.6 ° C " alternately.
Push once again the
- Probe tip temperature mode: the display is flashing and showing "ST"
and probe tip temperature e.g. "37.0 ° C " alternately.
Push once again the
first display ("CAL" and "
For leaving the CAL-mode wait approx.4 minutes, the appliance will turn off
automatically or push
39.7 ° C to 40.1 ° C
(103.5 ° F to 104.3 ° F ).
- button short time.
- button short time.
- button short time.
- button short time for reset to the
- - -
").
-button for > 3 seconds.