8 FAULT CHECKLIST

Fault
Bulb does not illuminate
Battery cannot be removed
All cuts are rough and ragged
Cross cuts are irregular
Scale of measuring
microscope blurred
Illumination of measuring
point not adequate
Cause
Battery flat
Bulb defective
Poor contact
Break in circuit
Unsuitable, non leakproof
battery used
Point of cutter tool damaged
Cross cutter worn out
Graticule scale holder loose
Bulb misaligned
Corrective Action
Insert new batteries
Insert new bulb
Remove slotted screw and scratch upper surface of
battery casing vigorously with coin until oxidated
metal is bare again
Return the gauge to your Elcometer supplier
Return the gauge to your Elcometer supplier
Replace tool
Replace tool
Return the gauge to your Elcometer supplier
Re-adjust bulb holder See "Replacing the bulb" on
page 4