14.4

Environmental Specifications

During Operation
During Transport
and Storage
Green Smart User Manual
To assure line voltage quality, a separate 3-core grounded power
cable connected directly to central distribution panel with an over-
current circuit breaker rated for 20A must be used.
Maximally allowed deviation of the tube voltage / tube current /
exposure time:
Tube Voltage (kVp) ± 10 % / Tube Current (mA) ± 20 % / Exposure
Time (s) ± (5 % + 50 ms) according to IEC 60601-2-63.
The mains resistance should not exceed 0.045 ohms at 100 V and
0.19 ohm at 240 V.
Item
Temperature
Relative humidity
Atmospheric pressure
Temperature
Relative humidity
Atmospheric pressure
14. Technical Specifications
Description
10 ~ 35 ℃
30 ~ 75 %
860 ~ 1060 hPa
-10 ~ 60 ℃
10 ~ 75 %
860 ~ 1060 hPa
145