Standard Calibration Adjust,
7. Press
ZERO/SCROLL
begin editing. The digit to be edited will flash. Use the
change the entry, then press
next digit.
When the last digit is entered, press
curve to the value entered. The meter will return to measurement mode and the
Calibration Adjusted icon will appear in the display window.
To turn off Standard Calibration Adjust (SCA):
1. Press
.
MENU
2. Press
ZERO/SCROLL
3. Press
READ/ENTER
4. Press
READ/ENTER
Note: Perform another standard calibration adjust to turn SCA on again.
Note: For meters with factory-calibrated ranges or methods, Standard Calibration Adjust
(SCA) will be disabled when a user-entered method is programmed into the meter. To
turn SCA back on, restore the meter to factory default calibration. See Retrieving the
Factory Calibration on page 2—25.
to access the Edit function, then press
READ/ENTER
until "SCA" appears in the display.
, then press
ZERO/SCROLL
to turn off SCA.
2—14
continued
to accept and advance to the
and the meter will adjust the
READ/ENTER
until "Off" appears in the display.
READ/ENTER
key to
ZERO/SCROLL
to