Function of LEDs

Connection
1
2
ye
3
gn
rd
4
A0023003
1: Level display
2: M12 plug
3: Valve plug
4: Cable
LED colors: gn = green, ye = yellow, rd = red
7.4

Function test with test magnet

WARNING
L
Danger of injury!
Ensure that no dangerous processes are triggered in the system.
Carry out a function test while the device is in operation.
Hold the test magnet for at least 2 s against the marking on the housing.
 This inverts the current switch status, and the yellow LED changes state. When the
magnet is removed, the switching status valid at that time is adopted.
If the test magnet is held against the marking for longer than 30 s, the red LED will flash: the
device returns automatically to the current switch status.
The test magnet is not included in the scope of delivery. It can optionally be ordered as
an accessory. See the "Accessories" -> "Additional accessories" section
Endress+Hauser
Maximum safety (MAX)
A0023004
A0023005
Operating modes
Minimum safety (MIN)
A0023007
A0023006
 not lit
  lit
 flashing
   fault/warning
Commissioning
Warning
Fault
A0023009