Test Setup

NPB-40
blood oxygen saturation. The test setup procedure sets up Test 2 for
75 percent blood oxygen saturation.
Test 3: Modulation — This procedure simulates an O
attached to a patient indicating low and high pulse strength. The test setup
procedure sets up Test 3 for low pulse strength.
Test 4: Light — This procedure simulates an O
patient indicating low and high light level passing through the patient at
the sensor site. The test setup procedure sets up Test 4 for low light level.
Note:
A DEC-4 extension cable may be used between the SRC-MAX and the
NPB-40.
1. Connect the SRC-MAX tester to the NPB-40 sensor port.
2. Turn on the NPB-40 by pressing the Power button.
3. After the NPB-40 completes POST, the NPB-40:
Is in SpO
alarm
2
Shows an %SpO
Shows a pulse rate of 60 (pass criteria is 57 to 63 bpm inclusive)
1 — Sensor Port
of 75 (pass criteria is 73 to 77 %SpO
2
Performance Verification
M
sensor
XI
AX
M
sensor attached to a
XI
AX
inclusive)
2
25