ABB Nitrate 8236 Instruction Manual page 20

Ise nitrate monitor
Table of Contents

Advertisement

...6 PROGRAMMING
...6.2.5 Current Output Page
Continued from previous page.
OP2 Cal Hold
Enter
OP2 Law
Enter
OP2 FSD
– – –
or
Enter
Lin
Log
OP2 Zero
or
Enter
18
Output 2 Calibration Hold
Yes
Current Output 2 can be held during calibration, if required. Select 'YES' or 'NO'.
No
or
Store.
Advance to next parameter.
Output 2 Law
Log
Current Output 2 can be either logarithmic or linear. Select 'Log' or 'Lin'.
or
Lin
Store.
Advance to next parameter.
Output 2 Full Scale
Current Output 1 full scale range: 10 to 5000 mg l
i.e. minimum span is 1 decade.
Set the required concentration value for Current Output 2 full scale.
Store.
Advance to next parameter.
– –
Output 2 Zero
Set the required concentration value for Current Output 2 zero.
Note. If the Output 2 Law is linear (Lin), this parameter is omitted and the zero
current output is automatically set to '0'.
Store.
Advance to next parameter.
Continued on next page.
–1
as NO
or 2 to 1000 mg l
–1
3
as N,