...6 PROGRAMMING
...6.2.5 Current Output Page
Continued from previous page.
OP2 Cal Hold
Enter
OP2 Law
Enter
OP2 FSD
– – –
or
Enter
Lin
Log
OP2 Zero
or
Enter
18
Output 2 Calibration Hold
Yes
Current Output 2 can be held during calibration, if required. Select 'YES' or 'NO'.
No
or
Store.
Advance to next parameter.
Output 2 Law
Log
Current Output 2 can be either logarithmic or linear. Select 'Log' or 'Lin'.
or
Lin
Store.
Advance to next parameter.
Output 2 Full Scale
–
•
Current Output 1 full scale range: 10 to 5000 mg l
i.e. minimum span is 1 decade.
Set the required concentration value for Current Output 2 full scale.
Store.
Advance to next parameter.
–
– –
Output 2 Zero
•
Set the required concentration value for Current Output 2 zero.
Note. If the Output 2 Law is linear (Lin), this parameter is omitted and the zero
current output is automatically set to '0'.
Store.
Advance to next parameter.
Continued on next page.
–1
as NO
–
or 2 to 1000 mg l
–1
3
as N,