Operating - ABB Nitrate 8236 Instruction Manual

Ise nitrate monitor
Table of Contents

Advertisement

6.2.1 Operating Page 1
The values displayed in Operating Page 1 are for viewing only and cannot be altered in this page.
Nitrate
<Unit>
or
xx
Millivolts
or
xx
Control Temp
or
Slope
xxx
or
Date
DD:MM:YY
or
Time
HH:MM:SS
or
Next AutCal
DD:MM:YY
or
Last Cal
DD:MM:YY
or
or
SECURITY CODE -----
Measurement Units
The measurement units are displayed, e.g. Nitrate ppm.
Advance to next parameter.
Mode
Sensor Output
xmV
The sensor output is displayed in millivolts.
Advance to next parameter.
Mode
C
Control Temperature
x
The heater block control temperature is displayed in degrees Celsius.
Advance to next parameter.
Mode
Slope Check Value
%
x
The slope value should be between 70 and 110%. If the value is outside these
limits check the electrode.
Advance to next parameter.
Mode
Date
The date is displayed.
Advance to next parameter.
Mode
Time
The time is displayed.
Advance to next parameter.
Mode
Next Calibration Date
The date when the next automatic calibration is to be carried out is displayed. If the
automatic calibration is disabled, 'OFF' is displayed in place of the date.
Advance to next parameter.
Mode
Last Calibration Date
The date of the last automatic or manual calibration is displayed.
Return to top of Operating Page.
Mode
or
Advance to Security Page (Section 6.2.3).