Performance Specifications
Table 10. CARBON MONOXIDE (appropriate models only) –
Typical Performance Specifications
RANGE
RESOLUTION
REPRODUCIBILITY**
RESPONSE TIME
* See Table 5. Instrument specification, NOTE
** See Table 14. PID Typical Performance Specifications, NOTE
Table 11. CARBON MONOXIDE – Cross Reference Factors for SIRIUS
Calibration Using Calibration Cylinder (P/N 10053022)
NOTE:
Data is presented as the indicated output in ppm, which would result from the
application of 100 ppm of the test gas.
TEST GAS (100 PPM)
Carbon Monoxide (CO)
Hydrogen SuIfide (H
Sulfur Dioxide (SO
Nitrogen Dioxide (NO
Nitric Oxide (NO)
Chlorine (Cl
2
Ammonia (NH
Hydrogen Chloride (HCI)
Ethylene (C
2
Hydrogen Cyanide (HCN)
Methane (CH
Ethanol (EtOH)
Hydrogen (H
52
500 ppm CO
1 ppm CO for 5 to 500 ppm CO
± 5 ppm CO or 10% of reading, whichever is greater
0 to 150 ppm CO ±15% >150 ppm CO
(normal temperature range*)
±10 ppm CO or 20% of reading, whichever is greater
(extended temperature range*)
90% of final reading in 50 seconds with sample line
and probe (normal temperature range*)
EQUIVALENT PPM
S)
2
)
2
)
2
)
)
3
H
)
4
)
4
)
2
100 ± 9
4 ± 4
0 ± 1
2 ± 6
70 ± 10
1 ± 8
2 ± 4
3 ± 2
76 ± 9
0 ± 1
0 ± 0
0
70 ± 26