Table A-4: N-3000 Internally Corrected Error Codes - Continued
Error Code Explanation
154
UIF RTXC executive function failed.
156
UIF stack communication bus common code failed.
157
UIF state machine illegal transition or unknown state.
180
External Port Service internal error (RS232 handler).
184
UIF Data stream or Reply timeout.
225
SpO
can't allocate a resource.
2
226
SpO
stack overflow.
2
250
SpO
general watchdog reset.
2
252
SpO
memory corruption.
2
253
SpO
unexpected interrupt.
2
254
SpO
RTXC executive function failed.
2
256
SpO
communication bus common code failed.
2
257
SpO
state machine illegal transition or unknown state.
2
278
SpO
cannot get power-on defaults from EEPROM or they are
2
invalid (bad values or low-limit above high-limit).
283
Operating mode changed during internal reset.
285
SpO
Controller PCB failure.
2
286
SpO
PCB failure.
2
287
SpO
detected failure of other system component (UIF PCB,
2
UIF software, or communication problem).
288
SpO
module failure (cannot determine whether SpO
2
Controller PCB or SpO
PCB).
2
Appendix
2
A-5