Section 4: Configuration Mode, Service Mode, and Alarm Active Function
4.2.8.2
UIF Software Version Report
4.2.8.3
SpO
Software Version Report
2
4.2.8.4
Set Serial Port Baud Rate
4.2.8.5
Trend Type
4-6
1.
From the configuration menu steady state, rotate the knob until "1" is
displayed in the SpO
button.
2.
The UIF software version number is the left-most digit in the SpO
display. The next two digits in the SpO
software revision number. The minor software revision number is
displayed in the PULSE RATE display. Press the LOWER ALARM
LIMIT BUTTON to return to the configuration menu steady state.
1.
From the configuration menu steady state, rotate the knob until "2" is
displayed in the SpO
button.
2.
The SpO
software version number is the left-most digit in the SpO
2
display. The next two digits in the SpO
software revision number. The minor software revision number is
displayed in the PULSE RATE display. Press the LOWER ALARM
LIMIT BUTTON to return to the configuration menu steady state.
1.
From the configuration menu steady state, rotate the knob until "3" is
displayed in the SpO
button.
Note:
Baud rates should not exceed 19,200 in RS-232 mode (use EIA-
422 mode). Refer to paragraph 6.5.1 to change the
communication mode settings.
2.
The current baud rate, in thousands, is displayed in the PULSE RATE
display. To change the default baud rate setting, rotate the knob until
the desired setting is displayed.
3.
Press the LOWER ALARM LIMIT button to return to the configuration
menu steady state.
Menu item number 4 allows you to select "5", "10" or "20" (displayed in the
SpO
% display) as the default trend format. Selecting "20" causes patient
2
trend data to be recorded every 20 seconds. Each patient parameter value
will be both the maximum and minimum data for each parameter during the
sample period. Data is stored for the most recent 24 hours of patient
monitoring. This selection is described as "Format 2" in the N-3000 operator's
manual.
Selecting "10" causes patient trend data to be recorded every 10 seconds.
Each patient parameter value will be the average of all data samples for each
parameter during the sample period. Data is stored for the most recent 24
hours of patient monitoring. This selection is described as "Format 1" in the
N-3000 operator's manual.
% display. Press the UPPER ALARM LIMIT
2
% display represent the major
2
% display. Press the UPPER ALARM LIMIT
2
% display represent the major
2
% display. Press the UPPER ALARM LIMIT
2
%
2
%
2