Section 4: Configuration Mode, Service Mode, and Alarm Active Function
4-8
4.
Press the PRINT button. You must press this button within 15 seconds
or the monitor will turn off automatically. You are now in service mode
steady state.
Note:
Failure errors (refer to Troubleshooting section for an
explanation of failure errors and error codes) may be
encountered by the N-3000 upon entering the service mode. The
N-3000 will automatically access the menu item used to correct
this situation.
If a user-correctable error code (a code beginning with "0") is
displayed while in the service mode, press the LOWER ALARM
LIMIT button to clear the error.
"SEr" stops flashing and is continuously displayed.
The number "1" is indicated in the SpO
The PATIENT MOTION indicator is lit, indicating that you are in
the service mode steady state, with access to the main menu as
indicated in Table 4-3.
5.
Use the knob to move from one main menu item to the next.
While in service mode, the UPPER and LOWER ALARM LIMIT buttons
are used as enter and exit buttons, respectively. You must press the
UPPER ALARM LIMIT button to select a main menu item and move to
the submenu level.
6.
When you have scrolled to the desired menu item, press the UPPER
ALARM LIMIT button. The PATIENT MOTION indicator goes out and
the PULSE SEARCH indicator illuminates.
This indicates that you are now in a submenu of the selected main menu
item. Use the knob to move from one submenu item to the next. Not all
menu items have submenu selections.
7.
To return to the service mode steady state from a menu item, press the
LOWER ALARM LIMIT button. (Menu item 3, Button Test, is an
exception; it is exited by rotating the knob.)
8.
To exit the service mode, power-down the monitor by pressing the
ON/STANDBY button. The N-3000 is automatically powered down if no
action is taken for approximately 5 minutes.
Main menu items available from the service mode steady state are discussed
in numerical order as indicated in Table 4-3.