Page 1
Hand-Arm Vibration Meter Type 2537 From serial no. 2002 521 August 1998 Brüel & Kjær BB1073 – 11...
Page 2
All rights reserved. No part of this publication may be reproduced or distributed in any form, or by any means, without prior consent in writing from Brüel &Kjær Sound and Vibration Measurement A/S, Nærum, Denmark. 0 – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 3
Basic introduction to the instrument: replacing batteries, using the buttons, reading the screen, and setting the language. Taking Measurements How to use the Type 2537: calibrating, taking a measurement, reading the display, and changing the set-up. Data Operations Saving and recalling measurement results, printing results, transferring results to a computer, using the AC output.
Chapter 8 — Index 1.1.2 Conventions Used in this Manual All references to buttons on the Type 2537 are shown with the button’s pictogram as it appears on the instrument. Refer to section 2.2 for a complete list of button pictograms and their functions.
Chapter 1 – Introduction Features Features The Type 2537 contains several features that enable you to take and save hand-arm vibration measurements under a variety of conditions. Features include: Ease of use Two measurement ranges Eight measured parameters (see below)
Settings can not be changed while the instrument is measuring. For more information about setting up the vibration meter for meas- urement, see section 3.5. 1 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
(at any time after a measurement is stopped). The records are stored in non-volatile memory, which is pow- ered by the backup battery. All records are preserved when you turn off the instrument. BB1073– 11 Hand-Arm Vibration Meter Type 2537 1 – 5 User Manual...
Erase all records For more information about records, see section 4.1. Backlight The Type 2537 includes a backlight, which makes the display easier to read in low light situations. Press to turn it on or off. To save batteries, the light will switch off automatically after 30 seconds.
Page 11
Replacing Batteries ............2 – 4 The Back-up Battery ............2 – 5 Pushkey Definitions ............2 – 6 The Measurement Window ..........2 – 8 Choosing the Language ..........2 – 10 BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 1 User Manual...
(at room temperature). When the indicator is blinking, there is less than one hour left. Under extreme cold or hot environments, the remaining time may be less. 2 – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 13
For long-term storage, remove the batteries and keep the vibration meter in a dry place. Never mix different makes or types of battery. Never mix charged and discharged batteries. BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 3 User Manual...
+/- orientation shown on the bot- tom of the battery compartment. Press the compartment lid back into place. Fig.2.3 The Type 2537 with the battery lid removed 2 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær...
If this happens, then your instrument’s calibration may no longer be valid. Contact your Brüel & Kjær dealer to have your hand-arm vibration meter recalibrated. BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 5 User Manual...
With these windows you can store, recall, erase, print and/or download records of your measurements. Range Press this key to step through the available measurement ranges. 2 – 6 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 17
Measure- ment Window. Backlight Switches on or off the display’s back- light. To save batteries, the back- light switches off automatically after 30 seconds. BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 7 User Manual...
(this is called a “latched overload”). To change the measurement range, press 2 – 8 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 19
L: Linear (6.3 Hz to 5 kHz) 10. RMS parameter: Shows the type of RMS reading dis- played (A , or Inst). Press to select the displayed parameters. 11. Value of the RMS parameter BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 9 User Manual...
Chapter 2 – Getting Started Choosing the Language Choosing the Language The Type 2537 is fluent in five languages. Use the procedure below to choose the language used in the instrument’s display screen. If the vibration meter is on, then press to switch it off.
Page 21
This means that your factory calibration is no longer valid and must be redone at a Brüel & Kjær service centre. Contact your Brüel & Kjær dealer for assistance. BB1073– 11 Hand-Arm Vibration Meter Type 2537 2 – 11 User Manual...
Page 22
2 – 12 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 23
When to Calibrate ............. 3 – 2 Principle of Calibration ............. 3 – 2 Which Calibrators Can I Use? .......... 3 – 2 Calibrating the 2537 Controller ........3 – 3 Mounting the Accelerometer .......... 3 – 4 Taking Measurements ............
Chapter 3 – Taking Measurements Setting the Measurement Time........3 – 16 Setting the Time and Date..........3 – 17 3 – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
The vibration meter can be calibrated with Brüel & Kjær Cal- ibration Exciter Type 4294 or any other certified vibration meter calibrator that emits a 159.2 Hz signal at 10 m/s BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 3 User Manual...
Page 26
If an error message appears after you press , make sure that: the calibrator is switched on the accelerometer is correctly and securely attached to the calibrator. the calibrator’s batteries are fully charged 3 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Brüel & Kjær dealer. Also, see section 5.4. Mounting the Accelerometer The Type 2537 comes with a mounting bracket, which enables you to attach the accelerometer to the tool or machine you are measuring. It enables you to take readings along all three axes (x, y, and z).
(check the measurement range on the screen as shown in Fig. 3.2). Begin a new measurement by pressing and repeat from step 4. 3 – 6 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 29
If there are no records left, then you will see a warning on your screen (see section 4.1.2 for more information). BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 7 User Manual...
Fig. 3.2). You may want to erase the last record of the under ranged measurement as de- scribed in section 4.1.4. Begin a new measurement by pressing and repeat from step 4. 3 – 8 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
(as described above). Thereafter, the upper limit of the measurement range will blink until a new measurement is started. BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 9 User Manual...
Page 32
Spreadsheet results have a special overload column in which overloads are indicated with “ ”, for yes (or the equivalent for language settings other than English). 3 – 10 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
The maximum RMS acceleration that occurred dur- ing the measurement period. The minimum RMS acceleration that occurred dur- ing the measurement period. The four-hour energy equivalent RMS acceleration. BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 11 User Manual...
Hand-held tools produce vibrations chiefly in the hands and arms of their users, which are most sensitive to vibrations in the range of 12 Hz to 16 Hz. The Type 2537 includes a hand-arm vibration weighting, which makes the instrument most sensitive to vibrations in this frequency range.
In some cases, you may want to use a linear setting. You may need this to conform with some local regulations or for use in research literature (as a supplement). The Type 2537 in- cludes a linear setting, which removes the hand-arm vibration filter.
If the display ever becomes full and the value for the upper limit of the measurement range changes to read “ ”, 3 – 14 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
To store the new frequency weighting, press . (If you prefer to cancel your changes and return to the old fre- quency weighting, press .) You will then return to the Measurement Window. BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 15 User Manual...
Page 38
Making Measurement Settings 3.5.3 Setting the Measurement Time The Type 2537 can be set to measure for a fixed amount of time. When the set time has elapsed, measurement is stopped automatically and the results are stored in the next available record (except when the off (manual) setting is used).
Page 39
The Measurement Window must read (as shown STOP in Fig. 3.8), not Press until the Date/Time Set-up Window is visible on the screen, as shown in Fig. 3.10. BB1073– 11 Hand-Arm Vibration Meter Type 2537 3 – 17 User Manual...
Page 40
To store the new date and time, press . It is not possible to cancel Date/Time changes. You will then return to the Measurement Window. 3 – 18 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 41
Transferring Results with Third Party Software ..4 – 16 Using the AC Output ............4 – 17 To Record Measurements to a DAT........ 4 – 17 BB1073 – 11 Hand-Arm Vibration Meter Type 2537 4 – 1 User Manual...
Working with Records Working with Records The Type 2537 is capable of saving up to 40 records of meas- ured results. This enables you to take several readings at different locations or at different times of day and store each one for future reference.
Page 43
4). You will then return to the Measurement Window. If you wish to cancel the operation rather than store the record, press return to the Measurement Window. BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 3 User Manual...
Always erase all records after you have downloaded and saved them to your computer’s hard disk. Erase records of invalid measurements (e.g. overloads) right away by using the Erase Last Record command (see section 4.1.4). 4 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Fig. 4.5. Note that if an overload occurred during the measurement, then the value for each parameter will blink. BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 5 User Manual...
Page 46
. To view later records, or to wrap back around to the first record, press When you have finished viewing your records, press to return to the Measurement Window. 4 – 6 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
. If you select to erase the record, then a warning like the one shown in Fig. 4.8 will appear to make sure you have not made a mistake. BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 7 User Manual...
Page 48
Fig. 4.11 will appear to make sure you have not made a mistake. 4 – 8 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 49
Fig.4.11 The Erase All Records Warning If you mean to erase all records, press again. Oth- erwise, press to leave all records stored and return to the Measurement Window. BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 9 User Manual...
Chapter 4 – Data Operations Using a Portable Printer Using a Portable Printer The Type 2537 includes a standard serial port interface. It is designed to work with portable serial printers. Contact your Brüel & Kjær dealer for more information about printers recommended for use with the Type 2537.
Page 51
If you have not taken a measurement, then take a com- plete measurement as described in section 3.3. Make sure measurement has stopped before proceeding to the next step. BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 11 User Manual...
Fig.4.13 The Print Measurement Window Press to begin printing. Press to cancel the operation. 4.2.2 Reading the Printout The Type 2537 generates a printout that supplies the follow- ing information: Frequency weighting Measurement range Date and time Elapsed time Overload status Sample printouts are shown in Fig.
Page 53
2.64 m/ss Fig.4.14 Print examples. In the example on the right an overload has occurred, which is indicated by the “Overload” label on the eighth line from the bottom BB1073– 11 Hand-Arm Vibration Meter Type 2537 4 – 13 User Manual...
Transferring Records to a Computer Transferring Records to a Computer The Type 2537 is capable of saving up to 40 records of meas- ured results. For more information about saving records, refer to section 4.1. All records can be downloaded to a computer in a standard spreadsheet format.
If Brüel & Kjær communication software is not installed on your computer, then locate the diskette and follow the instructions given on the label. Connect the Type 2537 to your computer as described above. Start running the Brüel & Kjær communication software.
DOS-only machine), you may want to use communications software other than that included with your instrument. If so, follow the instructions below. Connect the Type 2537 to your computer as described in section 4.3.1. Start your communications software. Set your software’s communication parameters as follows:...
Refer to your program’s user manual for instructions. Using the AC Output The Type 2537 includes an AC output port, which transmits vibration signals straight from the accelerometer’s input am- plifier. This is useful, for example, for recording the compete vibration signal to a DAT so you can bring samples back to a laboratory for further analysis.
Page 58
Begin recording on the DAT and use your vibration meter as usual. 10. If you intend to use an analyzer on the recording, then repeat step 8. 4 – 18 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 59
Storing the Instrument ............. 5 – 2 Cleaning the Instrument ..........5 – 2 Handling the instrument ..........5 – 3 Troubleshooting ..............5 – 3 Warnings and Error messages ........5 – 4 BB1073– 11 Hand-Arm Vibration Meter Type 2537 5 – 1 User Manual...
Chapter 5 – Maintenance and Troubleshooting Service and Repair Service and Repair The Type 2537 is designed and constructed to provide many years of reliable operation. However, if a fault occurs that impairs the vibration meter’s correct function, then remove the batteries to prevent risk of further damage.
Please contact your local Brüel & Kjær dealer for assistance. For help with error messages or other warnings that appear on the screen, refer to section 5.4. BB1073– 11 Hand-Arm Vibration Meter Type 2537 5 – 3 User Manual...
Chapter 5 – Maintenance and Troubleshooting Warnings and Error messages Warnings and Error messages If you get an error message while using or calibrating your Type 2537, then refer to the chart below. Message Occurs Solution Wait for measure- If you press...
Page 63
If you press when Make and store there are no records measurements *** WARNING **** saved and the buffer is before using this No Data empty. function. Available BB1073– 11 Hand-Arm Vibration Meter Type 2537 5 – 5 User Manual...
Page 64
5 – 6 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 65
Chapter 6 Specifications Specifications ................. 6 – 2 Ordering Information ............6 – 4 BB1073– 11 Hand-Arm Vibration Meter Type 2537 6 – 1 User Manual...
Page 66
1 s exponential averaging of the instan- taneous RMS readings (Inst). A is the highest peak reading (Peak) 0.01 100k Frequency (Hz) 960562e Fig.6.1 Frequency response with linear frequency weighting 6 – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 67
Size: 257 ×97 ×41 mm Factory set to CET. Weight: 460 g (incl. batteries) Warm-up Time: < 15s 0.01 0.001 0.0001 Frequency (Hz) 960561e Fig.6.2 Frequency response with hand-arm vibration weighting BB1073– 11 Hand-Arm Vibration Meter Type 2537 6 – 3 User Manual...
Page 68
AO 1386 9-pole Cable with 25-pole Adap- tor (for computer and serial printer) Brüel & Kjær reserves the right to change specifi- cations and accessories without notice. 6 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 69
Displayed Parameter Flowchart........7 – 5 Set-up Functions Flow Chart ......... 7 – 6 Measurement Range Flow Chart ......... 7 – 8 Data Functions Flow Chart ..........7 – 9 BB1073– 11 Hand-Arm Vibration Meter Type 2537 7 – 1 User Manual...
Battery level (see section 2.1.1) Run/Stop indicator Elapsed time Peak parameter (see section 7.2) Peak parameter value Frequency weighting 10. RMS parameter (see section 7.2) 11. Value of the RMS parameter 7 – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
The eight-hour energy equivalent RMS accelera- tion. Peak The maximum peak over the last 1 s interval. Inst A randomly sampled instantaneous RMS accelera- tion over the last 1 s interval. BB1073– 11 Hand-Arm Vibration Meter Type 2537 7 – 3 User Manual...
Measures with a linear (unweighted) response. The vi- bration meter will be equally sensitive to all frequencies from 6.3 Hz to 5 kHz. 7 – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
To Save Changes To Cancel Changes SET-UP Freq. Weighting From Fig. 7.4 >HA SET-UP Preset Time To Fig.7.4 >10 s Fig.7.3 Set-up functions flow chart, part 1 7 – 6 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 75
H Aeq4 ---- m/ss SET-UP Date/Time 04 Jan 1996 >22:15:02 SET-UP Date/Time 04 Jan 1996 22>15:02 SET-UP Date/Time 04 Jan 1996 To Fig.7.3 22:15>02 Fig.7.4 Set-up functions flow chart, part 2 BB1073– 11 Hand-Arm Vibration Meter Type 2537 7 – 7 User Manual...
Refer to section 3.5.1 for more information. STOP 00:00:00 ---- m/ss H Aeq4 ---- m/ss STOP 00:00:00 ---- m/ss H Aeq4 ---- m/ss Fig.7.5 Measurement Range flow chart 7 – 8 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
If there are no stored records and the buffer is empty, then the following warning message will appear when you press *** WARNING **** No Data Available Fig.7.6 The no data available warning BB1073– 11 Hand-Arm Vibration Meter Type 2537 7 – 9 User Manual...
Page 78
Current Measurement to Printer OUTPUT Stored Records (2x) To Erase Records Spreadsheet To Cancel ERASE Record No. n ERASE Records Fig.7.7 Data functions flow chart, part 1 7 – 10 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 79
10 Jun 15:00:45 the next Elapsed 01:00:00 parameter of this record H Aeq 7.26 m/ss STOP00:00:00 Amp---- m/ss H Aeq4---- m/ss Fig.7.8 Data functions flow chart, part 2 BB1073– 11 Hand-Arm Vibration Meter Type 2537 7 – 11 User Manual...
Page 80
7 – 12 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 81
Calibration ............ 3 – 3 choosing a calibrator ....... 3 – 3 range values .......... 3 – 10 errors ..........3 – 4 5 – 4 recalled parameters ......3 – 10 BB1073– 11 Hand-Arm Vibration Meter Type 2537 Index – 1 User Manual...
Page 82
Aeq ..........3 –11 7 – 3 Aeq4 ..........3 –11 7 – 3 Third party software ........4 – 16 Aeq8 ..........3 –12 7 – 3 Troubleshooting ..........5 –3 Index – 2 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...
Page 83
Index Under range ......... 3 – 7 3 – 8 Warning messages ........5 – 4 Weighting ............2 – 9 Weightings ............ 7 – 4 BB1073– 11 Hand-Arm Vibration Meter Type 2537 Index – 3 User Manual...
Page 84
Index – 4 Hand-Arm Vibration Meter Type 2537 Brüel & Kjær User Manual...