This instruction manual is for the microscope STM -3020B/T/BD . To insure safety and obtain optimum
performance and familiarize yourself fully with the use of this microscope .We recommend that you read
the manual thoroughly before operating the microscope ,Attain this manual instruction in an easily
accessible place near the microscope for the further reference.
Instruction Manual
Stereo Microscope
STM-3020 Series

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the STM-3020 Series and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Stalwart STM-3020 Series

  • Page 1 Instruction Manual Stereo Microscope STM-3020 Series This instruction manual is for the microscope STM -3020B/T/BD . To insure safety and obtain optimum performance and familiarize yourself fully with the use of this microscope .We recommend that you read the manual thoroughly before operating the microscope ,Attain this manual instruction in an easily...
  • Page 2: Table Of Contents

    CONTENT ................2 USER NOTICE 1. Components ................4 2. Assembly ..................5 2.1 Assembly Diagram ................5 2.2 Detailed Assembly Procedure ............6 3. Adjustment and Operation ............8 3.1 Adjusting Binocular Viewing Head ..........8 3.2 Focus Adjustment ................9 4.
  • Page 3: User Notice

    USER NOTICE 1. Safety Precaution 1. Carefully open the box, avoid the accessories, like lens, dropping to ground and being damaged. 2. Check the input voltage, be sure the input voltage which signed on the microscope is consistent with the power supply voltage, or it will bring a serious damage to the instrument. 3.
  • Page 4 the moisture. Do not disassemble any parts of the microscope, which will affect the function or decline the performance of the microscope. Place the instrument in a cool, dry position. When not using the microscope, keep it covered with a dust cover.
  • Page 5: Components

    1. Components Binocular head and zoom body WF10×eyepieces Diopter ring stand Clamping knob Focusing knob Standard base Power cord Glass stage Brightness adjusting knob Stereo Microscope STM-3020...
  • Page 6: Assembly

    2. Assembly 2.1 Assembly Diagram The diagram below shows how to assemble the various modules. The numbers indicate the order of assembly. ★When assembling the microscope, make sure that all parts are free of dust and dirt, and avoid scratching any parts or touching glass surfaces. ③...
  • Page 7: Detailed Assembly Procedure

    2.2 Detailed Assembly Procedure 2.2.1 Mounting binocular head (Fig.1) Loose the clamping bolt③ on the pillar④ and insert the binocular head and zoom body assembly②into the bracket of the pillar, then ① screw down the bolt③. ② ③ ④ 2.2.2 Mounting eyepieces (Fig.1、Fig.2、 Fig.3) Remove the eyepiece dust caps①...
  • Page 8 2.2.3 Connecting the power cord (Fig.4) 1. Clockwise rotate the left and right brightness ③ adjusting switches① until the end on (off) state before connecting the power cord②. 2. Insert the power plug② into the power jack of ① the microscope; make sure the connection is ②...
  • Page 9: Adjustment And Operation

    3. Adjustment and Operation 3.1 Adjusting Binocular Viewing Head 3.1.1 Adjusting interpupillary distance (Fig.7) Different users have different interpupillary ① distances. So when users change, do the interpupillary distance adjustment. While looking through the eyepieces, hold the left and right of the binocular assembly①and adjust the eyepieces by opening or closing them for binocular vision until the left and right fields of view coincide completely.
  • Page 10: Focus Adjustment

    3. 2 Focus Adjustment 3.2.1 Adjusting the rotation tension of the focus adjustment knob (fig.9、fig.10) 1. Rotate the focus adjustment knob① to move the slider② to a lower place for fear that the zoom ① body drops down suddenly if the tension is ②...
  • Page 11: Specification

    4. Specification 1:4.5 Zoom ratio: 1×~4.5× Objective magnification: Binocular viewing head: Inclined at 45, interpupillary distance range: 55~75mm 4.4 Working distance: 97mm Illumination system: Adjustable illumination for both LEDs (top and bottom) Eyepiece Working 10× 15× 20× Auxiliary distance Field of view 20 Field of view 16 Field of view 12 objective...
  • Page 12: Outfit

    3.The visibility of the Dust on objective front Clean lens surface image is poor. lens. Diopter ring is not Adjust it correctly. 4. Specimen image blurs correctly adjusted. when zoom magnification is Focus specimen Not in complete focus on changed. correctly at a high specimen.