Technical documentation Last changed on: 2020-12-02 EPS series Versatile Built-in or System Capable High Voltage Power Supply Module Versions from 500V – 30 kV • 60 W and 150 W versions • patented resonance converter technology • high efficiency • available as metal-box or 3U MMC (only 60 W) version •...
Advices marked as “Caution!” describe actions to avoid possible damages to property. CAUTION! INFORMATION Advices marked as “Information” give important information. INFORMATION Read the manual. Attention high voltage! Important information. EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 3/36...
You may only put the product into operation after it has been established that the high-voltage device complies with • the country-specific regulations, safety regulations and standards of the application. The high-voltage power supply unit may only be installed by qualified personnel. • EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 4/36...
WARNING! WARNING! Do not operate the unit in wet or damp conditions. WARNING! WARNING! Do not operate the unit in an explosive atmosphere. WARNING! EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 5/36...
Page 6
CAUTION! CAUTION! The devices (3UC) must only be used in combination with iseg approved crates. CAUTION! INFORMATION The shield of the high voltage output is always connected to the housing and can be used as a return conductor.
3.7 Troubleshooting 3.8 Maintenance 4 Dimensional drawings 5 Ground bolt connection 6 Connectors and PIN assignments 6.1 HV connectors 6.2 Interface connector D-SUB 9 (EPS metal box) EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 7/36...
Page 8
6.3 System connector H15 (EPS 3UC) 7 Order guides 8 Appendix 9 Glossary 10 Warranty & Service 11 Disposal 12 Manufacturer contact EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 8/36...
EPS modules can be equipped with ARC management or as capacitor charger with very low output voltage overshoot (option CLD). Customized versions can be produced on request. EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 9/36...
20% – 90%, non-condensing Notes: Specifications for stability, regulation, ripple and noise are guaranteed in the range 2% • V < V ≤ V Table 1: Technical data EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 10/36...
– monitor voltage, see 2.3 Options and Order Information z – monitor voltage item code (5V = 05, 10V = A0) ARC and 10V monitor not combinable Table 2: Configurations EPS 60W EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 11/36...
Page 12
– monitor voltage item code (5V = 05, 10V = A0) ARC and 10V monitor not combinable Table 3: Configurations EPS 60W with Option CLD EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 12/36...
– monitor voltage, see 2.3 Options and Order Information z – monitor voltage item code (5V = 05, 10V = A0) ARC and 10V monitor not combinable Table 5: Configurations EPS 150W CLD EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 13/36...
ARC management Capacitor charger Capacitor charger and ARC management VERY LOW NOISE Notes: for Set/Monitor voltage 5V only Table 6: Technical data: Options and order information EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 14/36...
(VLN – very low noise). • 3.1 Electrical wiring of the high voltage output Figure 1: Electrical wiring of the high voltage output EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 15/36...
To reset the error register, INHIBIT must be activated (set to low). EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 16/36...
400 ms after the supply voltage is applied and is then deactivated this function can optionally be disabled at the factory. • EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 17/36...
Regulation of the output current according to the set value of the output current under the condition measured output voltage (V ) < set value output voltage (V Figure 2: Operating area of the device EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 18/36...
The internal set value of the voltage is decreased from the set value to zero with the ramp speed depending on the configuration Switching off with INHIBIT: The HV generation is interrupted immediately, the set value of the voltage is set to zero. EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 19/36...
Table 8: Permissible number of ARCs. Device Number of ARCs / second EPS without option CLD EPS with option CLD Table 8: Permissible number of ARCs EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 20/36...
Table 9: Parameters of the ARC Management 3.4.6 VLN Reduced ripple see chapter 2 Technical Data. 3.4.7 CAR CAR is a combination the option ARC-Management and Capacitor Changer. EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 21/36...
5 V / (opt. 10 V) Notes: Case is connected to GND and 0V (Power supply ground) only without option ARC Table 10: PIN Assignment D-SUB 9 EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 22/36...
If KillEnable is active the reaching or exceeding of the I value will trigger a Kill-signal. This signal will switch off the HV immediately without ramp. Table 11: PIN Assignment H15 connector EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 23/36...
” of the connector “AIO”) or until the set value of the output current is reached (pin “I ” of the connector “AIO”, see chapter 3.5 Analogue I/O interface (AIO)). optionally 0 – 10 V EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 24/36...
Page 25
Figure 4: Electrical wiring of the analogue and digital in- and outputs EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 25/36...
It must be connected to the earthing bolt (see Figure 10: Ground bolt connection) on the rear panel. CAUTION! EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 26/36...
Repair and maintenance may only be performed by trained and authorized personnel. INFORMATION For compliance of the specified accuracy of set and monitor signals, the unit has to be recalibrated once a year. INFORMATION EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 27/36...
4 Dimensional drawings Figure 5: Dimensional drawing EPS 60W Cable Figure 6: Dimensional drawing EPS 3U Cassette with SHV HV-connector ≤ 8kV EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 28/36...
Page 29
Figure 7: Dimensional drawing EPS 3U Cassette with GES HV-connector ≥10 kV EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 29/36...
Page 30
Figure 8: Dimensional drawing EPS 150W EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 30/36...
5 Ground bolt connection Figure 9: Detail drawing of Figure 8 Figure 10: Ground bolt connection EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 31/36...
G31_C02-LLL Notes: Length building examples: 10cm ➜ 0.1, 2.5m ➜ 2.5, 12m ➜ 012, 999m ➜ 999 Table 13: Item code parts for different configurations EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 33/36...
For more information please use the following download links: This document https://iseg-hv.com/download/DC_DC/EPS/iseg_manual_EPS_ en . pdf Iseg crates (3UC) https://iseg-hv.com/de/products/detail/MME Crates Iseg THQ https://iseg-hv.com/de/products/detail/THQ Application note Coming soon EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 34/36...
HV ON/OFF HV OFF/ON channel(s) high voltage low voltage signal ground Inhibit Polarity KILL KillEnable EPS High Voltage Power Supply series | Last changed on: 2020-12-02 | www.iseg-hv.com 35/36...