3 Display / LEDs

When switched to "On", the NFA will briefly show "bAtt" and the re-
maining battery life. When switched to "Log", the remaining
battery life is shown cyclically with the elapsed log time and name of
the log file. After some self-tests the measurement starts automatically.
Time indications are set up as follows:
„Xd.XX": X days and XX hours or
„XX.XX": XX hours and XX minutes.
„XX.XX": (with blinking dot): XX minutes und XX seconds.
Two days is the specified minimum logging time ("2d00"). The real ca-
pacity can be higher, in particular with new batteries.
Status LED
The "Status" LED indicates the current mode of operation
"Red" indicates special operating conditions. Do not switch off while
the Status LED is red!
"Yellow" means "measurement in progress, NO recording".
"Green" means "measurement in progress WITH recording",
Note: with the switch set to "Log.", the "Status"-LED is switched off after a short while,
in order to extend battery life.
Special LCD Display Messages
"Low Batt": Please recharge the battery a.s.a.p.
„----": Autocalibration in progress, processor busy, or the value is too
small to be measured reliably.
„CHAr": Instrument is being charged.
„FULL"means: battery fully loaded or SD memory card is full.
„nbAt":The processor cannot recognize the battery
(see chapter 9)
OR the 15 V AC adapter should be disconnected
(the batteries have
reached their maximum possible capacity. Storage at this capacity level is not advised)
„Sd-L": short for SD memory card is locked.
„Sd-F": short for SD memory card failure". Try re-starting the NFA or
re-formatting the memory card according to chapter 9, first paragraph.
If necessary, replace SD memory card (max. 16 GB).