Basics BASICS Disposal of the measuring instrument Disposal information This instruction manual is part of the MultiPanel 2025 Do not dispose of this product with your instrument. It contains the information necessary for household waste. safe use. Read this manual before starting up the in- You can find more detailed information on strument.
Display Menu key Function keys On/off button (F4) The MultiPanel 2025 is a measuring instrument which provides the user with all functions needed for profes- sional and challenging metrology. When using sensors with ISDS (intelligent sensor de- tection), the measuring instrument automatically iden-...
Description of the measuring instrument Plug connector 1 Label Function U/F1 Analog signal input 10V Direction detection f-signal from channel 1 U/F2 Analog signal input 10V Direction detection f-signal from channel 2 Power supply sensor 1 (+Ub or 14VDC) Power supply sensor 2 (+Ub or 14VDC) Label Function...
Opens the menu for deleting measure- ment sere. Switches the view of the selected meas- urement or special channel on or off. Displays USB status: • Instrument connected with USB. • Instrument is communicating via USB. Displays battery status: Full, half-full, empty. MultiPanel 2025...
Start-up START-UP The display shows the firmware version number and serial number of the measuring instrument. Basic settings The display then shows the calibration notice. When you switch on the measuring instrument for the first time, you are asked to define the basic settings. How to define the basic settings [F4] Press...
Set the time format and time: [F1] format time • Use to select and press [MENU] [F1] [F2] • Press to change the selected value. [MENU] • Press to save your entry. [F4] DEVICE Press to return to the menu. MultiPanel 2025...
Calibration interval [F1] Calibration CALIBRATION INTERVAL Press to select the entry and [MENU] CALIBRATION to open the menu. The measuring instrument was calibrated before it was shipped by the manufacturer. The calibration in- terval is the period of time after which the measuring instrument is to be re-calibrated by the manufacturer.
PC or notebook. Follow the instructions displayed there. The installation file is downloaded. Run the installation file. Confirm the confirmation prompts, if applicable. Follow the installation instructions. The installation is complete when the link is dis- played on the desktop. MultiPanel 2025...
HYDROcenter Firmware update Note Instrument damage during firmware update If there is new firmware for your measuring instru- Errors during the firmware updates can damage ment, HYDROcenter offers you an update. the measuring instrument. Save all stored measurement series (e.g. with •...
• A zero adjustment cannot be executed for chan- nels showing a frequency. • No zero adjustment can be done for the following channels: C4, C5, C6, C7, C8. • Measurement values are set to zero for channels configured as counters. MultiPanel 2025...
Measurement [F3] Min./max. Display Start the measurement series by pressing [F1] Pressing switches to min./max. display or back The display shows a progress bar on the right to the current measurement values. side. [F2] Press to display the current memory status. Maximum values Minimum values [F2]...
[MENU] MAIN MENU Press to open the ment values again. [MENU] CHANNELS to open the menu. If you have switched on the display view for no channel or several channels, the following error is displayed: MultiPanel 2025...
Channels Configuring measurement channels C1 and C2 Measurement variable Unit of (measurand) measurement If you are using a sensor without ISDS, you have to Speed [mm/s] configure a channel. You need a corresponding pa- Angular speed [°/s] rameter for the sensor. They are on the sensor or in Ground [to] the sensor data sheet.
Page 16
Press until the display shows the measure- Configure the measurement variables: ment values again. [MENU] • Press [F1] [F2] • Press to change the measure- ment variable and unit of measurement. [MENU] • Use to save your entry. MultiPanel 2025...
Display Meaning of trigger input Inverse logic of the trigger input The trigger input of the MultiPanel 2025 is not, as for other measuring instruments from Hydrotech- nik GmbH; instead, it has an inverse logic. The trigger input is switched on ( ) if nothing is connected to the input.
[MENU] • Press to save your entry. [F4] The following values can be shown on the display: Press until the display shows the measure- ment values again. Display Meaning of trigger output Display Meaning of relay output MultiPanel 2025...
Channels Configuring analog outputs C7 and C8 Configure the output range: [F1] Range • Press to select [MENU] The operation of the menus for trigger output ( • Press [F1] [F2] and relay output ( ) is identical. • Press to change the first digit.
[MENU] Press to save your entry. [F4] DEVICE Press to leave the menu. [F1] Press three times to display the second page DEVICE of the menu. [F4] DEVICE Press to leave the menu. MultiPanel 2025...
Instrument configuration Units How to change the display rate DEVICE [F1] Open the menu and use to select Unit Disp.rate Use the setting to change the unit system of the entry. How to open the DEVICE menu, page 20 measuring instrument.
Error code Information text Date of last calibration Date [F4] You must confirm the information with in order to Read the following section for information on the DATE MENU use the measuring instrument. Time and date, page 8 MultiPanel 2025...
Resetting menu language RESETTING MENU LANGUAGE RESETTING MEASURING INSTRUMENT If the measuring instrument is set to the incorrect lan- You can reset the measuring instrument. Resetting guage, you may not be able to operate it. has the following effects: CHANNELS You can reset the menu language without under- •...
Error codes ERROR CODES TECHNICAL DATA Code Description Property Value 8100 Error in flash memory during read and Dimensions (H x W x D) 96 x 96 x 60mm write test Weight 298 g 8101 Flash memory could not be deleted Temperature range -20...+50 °C 8102...
Need help?