Ecg Settings; Setting The Lead Type; Setting The Lead Name - Fazzini FM5000C User Manual

Maternal/fetal monitor
Table of Contents

Advertisement

In order to obtain 1 mV calibration ECG waveform,ECG calibration should be conducted. During ECG
calibration, the screen displays the status message "CAL,CAN'T MONITOR!"(Cannot monitor the patient
during calibration).

15.6 ECG Settings

15.6.1 Setting the Lead Type

This monitor is applicable to the 3-lead, 5-lead and 12-lead ECG monitoring. When ECG cables of different
leads are used, the ECG leads that can be monitored are different.
When using 3-lead ECG lead wire, the leadscan be monitored include I, II and III.
When using 5-lead ECG lead wire, the leadscan be monitored include I, II, III, aVR, aVL, aVF and V1.
When using 12-lead ECG lead wire, the leadscan be monitored include I, II, III, aVR, aVL, aVF, V1 and V2,
and V3, V4, V5and V6.
When using the 3-lead ECG lead wire, only one channel ECG waveform can be displayed on the
screen; while when using the 5-lead or 12-lead ECG lead wire, only one ECG waveform can be displayed on
the fetal /mathor monitoring screen and big font screen, and ECG waveform of two channels can be
displayed on the mother monitoring screen, trend screen, list screen and oxyCRG screen.
Steps for setting the lead type:
(a) Click ECG parameter area and [ECG SETUP] menu will be displayed;
(b) Set the [LEAD TYPE] to [3 LEADS], [5 LEADS] or [12 LEADS].

15.6.2 Setting the Lead Name

(a) Click ECG waveform and [ECG WAVE] menu will be displayed;
(b) Click [LEAD NAME] and select the lead name from the pop-up list box, for example, lead II.
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
Figure 15-7Standard ECG waveform
15-8
User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fm5000plus

Table of Contents