Nibp Calibration - Fazzini FM5000C User Manual

Maternal/fetal monitor
Table of Contents

Advertisement

Patient
Monitoring
(3) Enter [NIBP SETUP] menu.
(4)
Select [PNEUMATIC] and the monitor starts to perform NIBP air leakage test with [PNEUM
TESTING...]status message displayedin the NIBP parameters area.
(5) System automatically inflates to the pressure of 180mmHg.
(6)
After about 20 seconds, the system will automatically open the valve, which means leakage
measurement is complete.
(7) If there is no error message in the NIBP parameter area, it means that the system has no sign of air
leakage. If error message [Pump Leaking...] is displayed, it means that the air circuit may have an
air leakage fault. Under such a circumstance, the operator should check the entire connection for
loosening, and run the leakage test again after having made sure that the connection is errorless.
(8) If the error message still appears, please contact the manufacturer for repair.
Warning
This air leakage test, different from those described in the standard EN 1060-1, is for users to
simply test air leakage in NIBP inflation. If the system displays an error message [Pump
Leaking...] at the end of testing, please contact the Our company maintenance engineers.

13.5 NIBP Calibration

Calibration of NIBP measurements should be carried out once a yearor when you doubt the measured NIBP
by the serviceman approved by Our company.
Manufacturers recommend a calibrated pressure gauge (or mercury sphygmomanometers) with a precision
more than 1mmHg to be used for calibration. Enter [Maintenance], enter the pasrword, and select [Calibration]
to begin calibration and this item becomes [Stop Calibration].
Pressure sensor calibration steps:
Metal containers with a volume of 500ml ± 5% instead of cuffs. Connect a calibrated standard gauge with
measurement error of less than 0.8mmHg, a spherical air pumps with a t-interface and inflatable tubes attached
NIBP
Pressure line
13-3
FM5000C - FM5000PLUS