The width of the cuff should be 40% (or 50% for neonates) of the circumference of the limb,
or 2/3 of the length of the upper arm. The length of the inflated part of the cuff should be long
enough to wrap 50~80% of the limb; a cuff of wrong size will produce a wrong reading. If you
have a problem about the size of the cuff, use a larger cuff in order to reduce errors.
Adult/pediatric/neonateresuable cuff:
Patient type
Limb Circumference
6~11cm
Neonate
10~19 cm
Infant
18~26 cm
Pediatric
25~35 cm
Adult 1
33~47 cm
Adult2
46~66 cm
Leg
Neonate disposable cuff:
Size
Limb Circumference
Neonate 1
Neonate 2
Neonate 3
Neonate 4
Neonate 5
Figure 18-2 Cuff usage
Cuff width
5cm
8 cm
10.6 cm
14 cm
17 cm
21 cm
3.0 ~ 5.5 cm
4.0 ~ 7.6 cm
5.6 ~ 10.6 cm
7.0 ~ 12.8 cm
8.9 ~ 15.0 cm
18-3
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
Gastube length
2 m
Cuff width
Gastube length
2.6cm
3.7 cm
4.5 cm
2 m
5.3 cm
6.0 cm