Setting The Alarm Delay Time; Pausing Alarms; Setting The Maximum Alarm Pause Time; Setting The Alarm Pause Time - Fazzini FM5000C User Manual

Maternal/fetal monitor
Table of Contents

Advertisement

8.5.4 Setting the Alarm Delay Time

The alarm system provides five kinds of alarm delay time: [DISABLED], [5SEC], [10SEC], [15SEC] and
[20SEC]. [DISABLED] indicates that the monitor will give alarms immediately once the measured
parameter value violated the alarm limit; [5SEC] indicates the monitor will give alarms only when the
measured parameter value exceeds the alarm limit for continual 5 seconds; and so forth. The alarm delay
function is not available on ECG parameter.
How to set the alarm delay time:
1.
Access [MAIN MENU], select [MAINTAIN] and enter the password;
2.
Select [ALARM SETUP]
3.
Select [ALM DELAY TIME] to adjust the alarm delay time.

8.6 Pausing Alarms

The users can press the
triggered and put the alarm system into alarm paused status.
1.
Stop all the indications of physiological alarms and no new physiological alarm will be triggered.
The remaining pause time "ALM PAUSE xxS" will appear at the physiological alarm message area.
2.
3.
[ALM PAUSE] on-screen key flashes in red.
4.
For technical alarm, the audio and visual alarm indicationwill be paused but alarm message still appears in
technical alarm message area. If new technical alarm is triggered during alarm pause, only the alarm
message of this technical alarm will be displayed without audio and visual alarm indication.
When the remaining alarm pause time ends, the alarm paused status will be automatically cancelled. You can
also cancel the alarm paused status by pressing the
on-screen key.

8.6.1 Setting the Maximum Alarm Pause Time

1.
Access [MAIN MENU], select [MAINTAIN] and enter the password;
2.
Select [ALM SETUP]
3.
Select [MAX ALM PAUSE TIME] to adjust the maximum alarm pause time. Available options are 2
minutes to 10 minutes.

8.6.2 Setting the Alarm Pause Time

1.
Access [MAIN MENU]
key on the front panel or click [ALM PAUSE] on-screen key to pause the alarm
FM5000C - FM5000PLUS
Maternal/Fetal Monitor
key on the keyboard or clicking [ALM PAUSE]
8-8
User Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fm5000plus

Table of Contents