Measurable wavelength of Masimo SpO2 module sensor: red light 660nm, infraredlight 940nm.
Measurable wavelength of Nellor SpO2 module sensor: red light 660nm, infraredlight 890nm.
The sensor's maximum output light power is less than 18mW.
Warning
This monitor can automatically identify the SpO2 sensor. However, as the internal hardware
has been fixed before delivery from factory, the use of improper SpO2 sensor will lead
to wrong measured values.

17.2 SpO2/PR Display

Pleth waveform

17.3 Safety Information

Warning
Check whether the sensor cable is in good condition prior to monitoring. When you
discennect the SpO2 sensor cable from the monitor, the monitor will trigger a technical alarm
and display alarm message "SpO2 NO SENSOR" on the screen.
If the sensor or its packaging has signs of damage, do not use it and return to the factory.
Continuous and prolonged monitoring may increase the risk of undesirable changes in skin
characteristics, such as irratation, reddenning, blistering, or even pressure necrosis etc,
especially for the neonates or patients with perfusion disorders and varying or immature forms
of skin. Check thesesor application site every two hours and move the sensor if the skin
quality changes.More frequent checkis required due to different status of individual patents.
High oxygen levels will have premature children been in danger of infecting the crystal-like
fibrous tissue disease. Therefore, the SpO2 alarm limit must be set carefully based on generally
accepted clinical practice.
Do not entangle the cable of the sensor and electrical surgical unit.
Do not placethe sensor on limbs with an arterial catheter or intravenous tube.
Do not place the sensor and the cuff on the same limb while measuring, because the NIBP