Hitachi UST-5550 Instruction Manual page 9

Intraoperative electronic linear probe, compatible with waterproof cover
Table of Contents

Advertisement

Warning
During surgery
, be sure to wear sterilized medical gloves.
Conducting examinations with the bare hands can expose the operator to a risk of infection.
For the acoustic medium, use sterilized physiological saline.
Using an unsterilized ultrasound medium can cause an infection on the patient.
Dispose the probe used for patients with Creutzfeldt-Jakob disease.
Otherwise, there is a risk of infection to the operator or patient. Our ultrasound probe is not compatible
with any disinfection/sterilization method for Creutzfeldt-Jakob disease.
When using ultrasound contrast agent, follow the supplied documentation.
Unexpected accidents could result. Check the state of the patient and take appropriate precautions
to avoid side effects.
Do not use the equipment fallen on to floor.
Ohterwise, there is a risk of infection. Stop the operation and perform the procedure in section 8
"Periodic Inspection", section 5 "Cleaning, disinfection and sterilization" and section 3-1 "Start up
check".
Caution
Constantly check for anything abnormal about the patient's condition and equipment.
Continued use without noticing that an abnormal condition has occurred can result in an electric
shock and injury to the operator or patient. If an abnormal condition occurs, immediately move the
equipment away from the patient and stop use of the equipment.
The equipment is vulnerable to damage by impact. Therefore, handle it with care.
There is a risk of damage to the equipment when the equipment is fallen or hit somewhere.
Do not use this probe with other equipment except for those specifically approved in the manual.
Use with unapproved equipment can result in an electric shock, burn, or other injury to the patient or
operator and damage to the probe and the other equipment.
Scan for the minimum length of time necessary for the diagnosis and at the lowest suitable output.
Overuse can adversely affect the internal tissues of the patient.
For details about the acoustic output, please refer to the documentation supplied with the ultrasound
diagnostic instrument.
Do not use it in continuous contact with the human body more than 60 minutes.
Overuse can adversely affect the internal tissues of the patient.
Regularly perform maintenance inspection and safety tests of the equipment.
If you use equipment for a long period of time, it can reduce the performance, or cause smoke or
fire. If anything unusual occurs, immediately stop using it and contact one of our offices and/or
distributor's offices listed on the back cover.
Use, move and transport the equipment under the environmental conditions specified in this
manual.
Otherwise, it may be damaged.
See section 2-5 "Environmental conditions" and section 7-4 "Environmental conditions during
transportation".