4.1. Select the Plus button to create a new stage. If the method will be a single stage method, the user
does not need to return to the Stages tab. For multi-stage methods, return to the Stages Tab after
setting the method parameters to create additional stages.
5. Select the "Parameters" tab.
5.1. Once in the system parameters, determine if the stage will be a Set Time or Constant Weight analysis.
NOTE
If creating multiple stages, only the final stage can be a Constant Weight stage. All previous stages must
use Set Time.
5.2. Set Time - The method will run for a preset length of time, regardless of sample size or final weight.
5.2.1. Set the Mag Power and IR power . The power should be set at 0%, 25%, 50%, 75%, or 100%
Method - Create, Edit and Delete
39