6.3

X-ray Exposure

1.
Get out of the X-ray room and close the door.
2.
Press and hold down the Exposure Switch until image acquisition is completed.
3.
Release the Exposure Switch when "Image capturing is completed" message
appears on the screen.
4.
Remove the Plaster Cast out of the equipment.
Green X (PHT-75CHS) User Manual
6. Acquiring 3D MODEL Scan Images
Do not operate the PC during exposure. Doing so may
cause the system to malfunction.
Yellow: X-ray On
The image appears on the screen.
During X-ray exposure, the status appears as follows.
The LED light of the Exposure Switch turns yellow.
The LED light on the top of the equipment turns yellow.
An alert sound comes out to indicate that X-ray emission
is currently underway.
On Console Software, the radiation mark turns Green.