Gamry Instruments Interface 1000 Operator's Manual page 48

Table of Contents

Advertisement

Chapter 6: Instrument Circuitry--Interface 1000 Schematic/Block Diagrams
Notes for Figure 6-3:
This diagram shows one of two identical ADC channels. One channel is dedicated to measurement of
the potentiostat's current signal and the other is used to measure the voltage signal.
Both A/D converters are triggered simultaneously to start a conversion. This trigger and the pulse
updating the Scan DAC voltage are under the control of a hardware state-machine. This insures that all
waveform and data-acquisition timing is tightly controlled and reproducible point-to-point.
By default, the data acquisition is synchronized with the system's 600 kHz power-supply switching-
frequency reducing noise from the power supply. Data-acquisition times that are a multiple of 1.666
µs will maintain this synchronization.
All analog signals are received differentially as shown here.
The 5 Hz, 1 kHz, 10 kHz, 200 kHz and 1.4 MHz filters are two-pole Butterworth filters.
The voltage-measurement channel includes a programmable 1×−0.25× gain circuit. In the 0.25 gain
position, voltages as high as 12 V can be measured.
Figure 6-3
One A/D Signal Chain in the Interface 1000
6 - 4