Download Print this page

Hach AS950 Basic Installation And Maintenance page 11

Hide thumbs Also See for AS950:

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
5.1 Site installation guidelines
Explosion hazard. The instrument is not approved for installation in hazardous locations.
Fire hazard. This product contains a flammable refrigerant. Do not damage or puncture the
refrigeration circuit. Make sure that all airflow openings in the instrument and in the structure (if
applicable) are not blocked.
• Install the AWRS in an indoor or outdoor location.
• Make sure that the temperature at the location is in the specification range. Refer to
on page 4.
• Install the sampler on a level surface. Adjust the sampler feet to make the sampler level. Refer to
Figure 2
on page 7 for the sampler dimensions.
• Use the installed anchor brackets and user-supplied
• Plumb a drain tube to the ½ in.-14 NPT female connector on the bottom of the sampler.
Figure 4 AWRS anchor bracket locations with mounting dimensions
1 Optional anchor brackets
5.2 Prepare the sampler
5.2.1 Clean the sample bottles
Clean the sample bottles and caps with a brush, water and a mild detergent. Flush the sample
bottles with fresh water followed by a distilled water rinse.
5.2.2 Install a single bottle
When a single bottle is used to collect one composite sample, do the steps that follow. When multiple
bottles are used, refer to
When the bottle is full, the full bottle shut-off stops the sampling program. Install the sample bottle as
shown in
Figure
5.
D A N G E R
W A R N I N G
Install multiple bottles
3
/
-in. bolts for the AWRS. Refer to
8
2 Anchor brackets (2x)
on page 12.
Specifications
Figure
4.
English 11

Advertisement

loading

This manual is also suitable for:

Awrs