pin 8 of the ADCs (U23, U22, U21, U20) should be at least
0.2 V above the lower limit and that the highest voltage pulse
on any pin 8 of the ADCs should be at least 0.3 V below the
upper limit. If an interpolator has a voltage pulse outside the
limits the measurement result will be wrong. Figure cc shows
the signals on an ADC.
The signal from the prescaler is connected to pin 22 (sin-
gle-ended) of U11. It comes via a level converter. Check the
input signal to the converter at R335 (PECL levels).
If the FPGA or a part in the interpolators has been changed or
repaired, a calibration of internals must be performed after-
wards. See Chapter 7.
Note
The interpolator design has varied slightly in the
course of time, but the pulses in Figures 6-49 and
6-50 are very little affected.
X3
Figure 6-48
Important locations for the measuring logic.
6-28 Troubleshooting
R623
X4
R335
U11
Figure 6-49
U23
U20
U21
U22
ADC 10461 behavior. Time A-A, 10 MHz in.
U20:8
U21:8
U22:8
U23:8