ENV4200
To activate remote control, apply one of the pins 14, 15, 16 or 17 to GND (pin 12).
5.3 Test receiver control menus
The independent R&S ENV4200 control menu is used for remote control with R&S EMI
test receiver types ESW, ESL, ESRP, ESR, ESU, and for EMC32 and ELEKTRA soft-
ware. The EZ-21 connecting cables (3 m or 10 m) are used for open-loop control with
the ESU test receiver. The EZ-29 connecting cables (3 m or 10 m) are used for the
ESL, ESRP, ESR and ESW test receivers.
For remote control with R&S EMI test receivers ESHS10, ESHS20/30, ESAI/ESBI/
ESMI, ESCS, ESS, ESPI, ESCI, ESIB and in the ES-K1 and ESXS-K1 EMI software,
the operating menu for the ESH2-Z5 V-network also applies for control of the
R&S ENV4200.
For control of the R&S ENV4200 with the R&S ESxI EMI test receivers ESAI; ESBI;
ESMI, the EZ-22 connecting cable is used. For control with other test receivers, the
EZ-21 connecting cables (3 m or 10 m) are used.
The pulse limiter cannot be remote controlled. For reasons of safety, the pulse limiter is
switched on during remote control. If impairment of the disturbance spectrum is sus-
pected due to pulse limiting, the pulse limiter can be switched off on the front panel and
the measurement repeated manually.