R&S ENV4200 User Manual page 22

Four line v-network
Table of Contents

Advertisement

®
R&S
ENV4200
4. Connect the black MC cable sockets of the supply voltage with the labels "L1", "L2"
and "L3" ('phase' or 'line') to the MC high current connections with the "L1", "L2"
and "L3" identifiers on the rear panel of the R&S ENV4200.
4.7 Example: Measurement setup
The following example shows a test setup for measuring disturbance voltages on a
hand drill. The artificial hand must be connected for this type of EUT. Three metal foils
are arranged on the hand drill and connected to the artificial hand.
1 = Metal wall, at least 2 m x 2 m
2 = Equipment under test
3 = AC supply cable
4 = Connection cable running separately to the hand simulation
5 = R&S ENV4200
6 = Metallic collar capable being gripped
7 = Test receiver
A = Power supply connection
B = Reference ground connection, low-inductance (e.g. brass plate, 0.2 mm)
P = Device under test connection (L1, N)
H = Artificial Hand connector
M = Test receiver connector
D = Metal foils
User Manual 1107.2470.12 ─ 04
Operating the R&S ENV4200
Example: Measurement setup
22