Agilent Technologies E1439 User Manual page 56

Vxi 70 mhz if adc with filters and memory
Table of Contents

Advertisement

Using the Agilent E1439
Fiber Optic Interface
Copy
The Copy fiber mode copies optical data from its fiber receiver to its fiber transmitter without
adding any data. Normal data collection and filtering continues, and the data port selection
determines whether data is sent to the local bus (Agilent E1439D only) or read from the FIFO via
the VME bus. Copy is the default fiber mode after power-on or reset. See the following
illustration.
Fiber Interface Setup
Fiber Mode
Rate
BOF
CRC
Flow Control
Epoch Generate
Epoch Size
Note
Setting the data port to Fiber while in the Copy fiber mode results in an invalid instrument state.
46
E1438D / E1439D
Fiber RX
DIGITAL
ADC
FILTERS
Copy
106 or 250 MBs
setting ignored
must match incoming signal
setting ignored
setting ignored
setting ignored
1 KB
Fiber TX
FIFO
FIFO
VME BUS
LBUS

Advertisement