Loop Slings 5 Safety Instructions 6 General 6 Shock Prevention 7 Equipment Warning Labels 7 Operating the Maxi 500 8 Power base models 8 Correct use of the brakes 8 Emergency shut-off 9 Raising and lowering a patient 9 Transferring a patient 10...
Box 61 S-241 21 Eslöv Sweden Service and support A service routine has to be performed on your Maxi 500 by ARJO trained personnel to ensure the safety and operation of your product. See chapter Care and Preventive Maintenance. If you require further information, please contact your local...
Operating and Product Care Instructions. left The Maxi 500 is intended to be used with clip slings or loop slings, depending on the spreader bar. Use only slings that are designed for Maxi 500.
Compatible Slings ARJO clip sling profiles that can be used with the Maxi 500. Clip slings Four-point padded sling Four-point amputee sling Four-point toileting sling (with headrest) Four-point unpadded sling Four-point mesh sling For more information about clip slings and how to use them, refer to the...
Compatible Slings ARJO loop sling profiles that can be used with the Maxi 500. Loop slings Hammock (THA-X) Hammock 6 (THA6-X) Quick-fit (TIR-X) Hygienic (THY-X) Combi 626002 Repositioning 624500 For more information about loop slings and how to use them, refer to the...
• Always carry out the recommended daily maintenance before using the lift. for the purpose shall be used for the Maxi 500, in order to • Always maneuver the lift with the handles provided. avoid injuries attributable to •...
Do not touch a damaged power cord. Electrically live equipment can result in serious injuries. If the lift or charger has any exposed or damaged wires, contact your local ARJO representative immediately. • Check nameplate for voltage and cycle require- ments.
Operating the Maxi 500 The Maxi 500 is delivered to you in a fully Power base models assembled state. Unpack the battery pack supplied, The base of the lift can be set with the legs opened or and fully charge it as described in the section closed by using the two bottom buttons on the hand “Battery Charging.”...
Operating the Maxi 500 Correct use of the brakes Raising and lowering a patient The electrical actuator can be operated in both Brakes application and release: directions and allows the operator to raise or lower the patient without any physical effort. The hand •...
WARNING: Emergency Lowering This mobile floor lift is not intended to be The Maxi 500 is equipped with a special safety a transport device. A patient in the lift feature that allows the boom to be lowered even if should not be moved more than a few the electrical system does not operate.
To ensure that the Maxi 500 is always ready for use, Insert the battery pack into the charger unit and it is recommended that a fully charged battery pack push it firmly.
The charger must not be covered or exposed to dust. The metal contacts must be kept clean. WARNING: • The battery charger is for use only with Maxi 500 batter- Ensure that the electric power ies that are supplied by ARJO. to the charger unit is switched •...
(on the edge of a bed, in a chair, wheelchair or similar). 13 Lower the patient and unfasten the sling from the spreader bar. Move the Maxi 500 away before removing the sling from under the patient. WARNING: When using clip slings (see pic-...
Position nect the leg strap clips/loops. the Maxi 500 so that the spreader bar is just above Adjustment of the spreader bar height may have and centered over the patient.
Digital Scale Model 1000 Scale Digital Scale Calibration The MODEL 1000 weight module is a compact and The serial numbers for display and load cell must precise scale system integrated into the patient lifts be identical. and specifically designed for patient lifts. Completely self-contained, the MODEL 1000 scale Remove the screws from the back of the display combines the latest precision strain gauge technology...
221.20265 POWER BASE SHIFTER PLATE 236.20801 MAXI 500 LEFT LEG 236.20806 MAXI 500 RIGHT LEG 236.20811 MAXI 500 BASE 203.20835 SCREW BASE MAXI 500 700.20131 KIT ROD END M10 297 LR 700.20880 MAXI 500 POWER BASE ACTUATOR 000.04430 FLAT WASHER M10 ZINC (20 OD)
Care and Preventive Maintenance Preventive maintenance schedule The Maxi 500 is subject to wear and tear, and the following actions must be performed when specified to ensure that the product remains within its original manufacturing specification. WARNING: The points on this checklist are the minimum the manufacturer recommends. In some cases more frequent inspections should be carried out.
Page 25
18. Ensure that the legs are parallel to each other with the help of a square. Make sure that the legs are perpendicular at 90 degrees to the base. If the product does not work as intended, immediately contact your local ARJO representative for support.
• Call for service (charger may be faulty). lights are visible. Yellow indicator light does not go • Internal batteries need replacing. Call Arjo for replacement. off after several hours of charging time. Battery pack indicates it is fully •...
1. Safe Working Load (S.W.L.) of 227 kg (500 lb) 2. Product name 3. Read operating instructions before using the product 4. ARJO logo 5. Emergency stop and reset button identification 6. Date of manufacture, product code, serial number and manufacturer...
Page 30
432 - 00133 ROMA Tel. 06 87426211 Fax. 06 87426222 Email: promo@arjo.it If your country is not listed here, please contact your local distributor, or Arjo International AB, Box 61, S-241 21 ESL Ö V, SWEDEN. Tel +46 413 64500. Fax + 46 413 555586.
Need help?
Do you have a question about the MAXI 500 and is the answer not in the manual?
Questions and answers