Locator Probe Function Check - EA Technology UTL2 Operating Manual

Ultratev locator probe
Table of Contents

Advertisement

EA Technology
2.3

Locator probe Function Check

The Locator probe function check is designed to allow users to verify the functionality of their
instrument using an in built test source. From the Main Menu, navigate to Function Check then
toggle on the Locator Test Source and navigate back to the Locator Advanced screen. Hold the
Locator probe and the UltraTEV Plus
will show the histogram peak on the left hand side of 0, closest to the Locator probe, as shown
below:
Function Check Menu
Do not take measurements with the test source turned on.
The Calibrate process and Function Check are not a substitute for annual factory
calibration.
UltraTEV Locator Probe Operating Manual
together (TEV plate to TEV plate), the Locator Advanced screen
2
Page 5 of 19
Function Check in Locator Advanced Screen
2852-MANUL-V01.03.00

Advertisement