READ ALL INSTRUCTIONS BEFORE OPERATING EQUIPMENT
Troubleshooting (continued)
Too Many Uses
Cause
The application has been dispensed by that particular
user to their max permission level within a 24-hour
period.
Warning: No Calibrations Near Dispense Size
Cause
There are no calibrations close to the amount of
chemical that is attempting to be dispensed.
Pump Needs Calibration
Cause
The application trying to dispense is time-based and
needs the pump to be time-calibrated, but the pump
is not.
Scale Not Assigned
Cause
The scale has not been assigned to a manifold in the
Admin screen.
Page 22 of 33
ALX-PRO Manual V5
Solution
1. Login to Clean Intel
2. Navigate to Users and edit the user's permissions
to be able to dispense that application more
times within a 24-hour period.
3. Sync the unit to push changes to the ALX-PRO.
Alternative: under General Settings select "Don't Use
Permissions" (See Page 8)."
Solution
1. Navigate to Pump Configuration and assign the
chemical to a device.
2. Select the pump requiring calibration
3. Calibrate a value close to the dispense amount.
(See Page 14-15).
Solution
1. Navigate to Pump Configuration.
2. Select the pump requiring calibration and select
Time Cal.
3. Set calibration points as necessary (See page 15).
Solution
Navigate to Manifold Configuration screen and assign
the scale to the manifold. (See Page 11).
Updated: 02/22/2022