C. Step By Step Procedure
Operation
1
Disassembling
2
Pre-Disinfection
3
Rinsing
4a
Automated cleaning with
washer-disinfector
6 .
MAP System . Directions for use
Operating mode
Disassemble the device, if required
Soak all instruments immediately after use in a
detergent and disinfecting solution combined
with proteolytic enzyme if possible.
Abundant rinsing (at least 1 min)
– Place the devices in a kit, box or container to
avoid any contact between instruments.
– Put them in the washer-disinfector (Ao value
> 3000 or, at least 5 min at 90 °C).
A1. Instruments
Warning
Silicone stops have to be removed.
– Follow instructions and observe concentrations and
immersion times given by the manufacturer (an excessive
concentration may cause corrosion or other defects on
instruments).
– The disinfecting solution should be aldehyde free (to avoid
blood impurities fixation) and without di- or triethanola-
mines as corrosion inhibitor.
– Do not use disinfecting solutions containing Phenol or any
products which are not compatible with the instruments
(see general recommendations).
– For visible impurities observed on instruments a pre-clea-
ning is recommended by brushing them manually with
soft material.
– Use quality water in accordance with local regulations.
– If a pre-disinfectant solution contains a corrosion inhibi-
tor, it is recommended to rinse the instrument just before
the cleaning.
– Discard any instruments with large obvious defects
(broken, bent).
– Avoid any contact between instruments. When placing
in the washer-disinfector use kits, supports or container.
– Follow instructions and observe concentrations given
by the manufacturer (also see general recommendations).
– Use only approved washer-disinfector according to EN ISO
15883, maintain and calibrate regularly.
Following
First use
uses