Clean the inside of the head using the cleaning curette. To clean the head thoroughly, insert the cleaning
curette with a twisting, clockwise movement. Pull it out, again with a twisting, clockwise movement.
12
Clean the parts with a cleaning and disinfecting product.
If necessary, clean the aluminum sterilization case (avoid using a dishwasher, ultrasonic bath and acid or
basic chemicals for cleaning).
Dry the parts with compressed air.
Attach one or more silicone rings to the head (refills are available in dispensers,
20253 red,
20254 blue and
ref
ref
Place the parts in the sterilization case, without inserting the pistons in the corresponding heads.
Sterilize in the autoclave or by hot air.
7. Disinfection, cleaning and sterilization
Reprocessing procedure for dental instruments and radicular devices. Do not exceed 200°C (392°F)
Foreword
For hygiene and sanitary safety purposes, all instruments not marked «sterile» must be cleaned, disinfected and
sterilized before each usage to prevent any contamination.
A. Area of application
Disinfection and Sterilization before first usage and reprocessing procedures concerning:
1. Instruments
Root canal filling instruments (Pluggers, spreaders, compactors).
Supports, kits and instrument organizers.
B. Filling material
Please refer to the individual Direction for use.
General recommendation
Use only a disinfecting solution which is approved for its efficacy (VAH/DGHM-listing, CE marking, FDA approval)
and in accordance with the DFU of the disinfecting solution manufacturer. For all metal instruments, it is
recommended to use anticorrosion disinfecting and cleaning agents.
For your own safety, please wear personal protective equipment (gloves, glasses, mask).
The user is responsible for the sterility of the product for the first cycle and each further usage as well as for the
usage of damaged or dirty instruments where applicable after sterility.
Limitations and restrictions on reprocessing: The appearance of defects such as cracks, deformations (bent,
twisted), corrosion, loss of color coding or marking, are indications that the devices are not able to fulfill the
intended use with the required safety level.
Single use marked instruments are not approved for re-use.
The water quality has to comply with the local regulations especially for the last rinsing step or with a washer-disinfector.
Plastic supports, hand instruments and NiTi instruments are degraded by Hydrogen Peroxide (H
NiTi Instruments are degraded if immerged more than 5 minutes in a solution of NaOCl at more than 5 %.
Instruments made of aluminum are degraded in presence of caustic soda solutions with mercury salt.
Do not use acid (pH < 6) or alkaline (pH > 8) solutions.
The washer-disinfector is not recommended for aluminum sterilization case.
20255 green).
ref
20252 yellow,
ref
O
) solution.
2
2