(12) Frequency
Range
100.00Hz
1000.0Hz
10.000kHz
Minimum Sensitivity :
> 5V (for ACV 1Hz ~ 10kHz)
> 8A (for ACA 1Hz ~ 1kHz)
Minimum Frequency : 1Hz
Overload Protection : AC/DC 1000V and 600A
(13) HFR (High Frequency Rejection)
Available for ACV, ACA and, Flexible Current Probe.
Add ± 4% to specified accuracy of each function and each range for 45Hz to 200Hz.
Accuracy is unspecified for > 200Hz.
Cut-off Frequency (-3dB) : 800Hz
(14) Inrush Current
Available for ACA and Flexible Current Probe.
Trigger level : ≧50d.
Add ± 3% to specified accuracy of each function and each range.
(15) Resistance
Range
600.0Ω
6.000kΩ
60.00kΩ
600.0kΩ
To get more accuracy result, short the test probes to obtain the offset. The accuracy
specification is specified for the result that the offset is subtracted.
Overload Protection : AC/DC 1000V
(16) Continuity
Built-in buzzer sounds when measured resistance is less than 20Ω and sounds off
when measured resistance is more than 200Ω, Between 20Ω to 200Ω the buzzer
maybe sound or off either.
Continuity Indicator : 2.7K Tone Buzzer
Response Time of Buzzer : < 100msec.
Overload Protection : AC/DC 1000V
(17) Diode
Range
1.500V
Open Circuit Voltage : Approx. 1.8V
Overload Protection : AC/DC 1000V.
OL Reading
100.00Hz
1000.0Hz
10.000kHz
OL Reading
660.0Ω
6.600kΩ
66.00kΩ
660.0kΩ
OL Reading
1.550V
16
Resolution
Accuracy
0.01Hz
0.1Hz
±(0.3% + 3D)
0.001kHz
Resolution
Accuracy
0.1Ω
±(0.9% + 5D)
0.001kΩ
0.00kΩ
±(0.9% + 2D)
0.1kΩ
Resolution
Accuracy
0.001V
±(0.9% + 2D)