Safety and Setup Guide
S1P-2, S1P-4, S1A-10, S1A-12, S1A-22
04-2021

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TRIOS S1P-2 and is the answer not in the manual?

Questions and answers

Summary of Contents for 3Shape TRIOS S1P-2

  • Page 1 3Shape TRIOS Safety and Setup Guide S1P-2, S1P-4, S1A-10, S1A-12, S1A-22 04-2021...
  • Page 2: Table Of Contents

    To learn more about 3Shape, our products and global office locations please visit our website: www.3shape.com Date of manufacture If you would like to receive a free printed copy please contact 3Shape support via email,  support@3shape.com. Manufacturer General warning sign– (black on yellow)
  • Page 3: Description Of Trios

    Type BF applied part Identifies the calibration tip Medical Device WEEE Product disposal symbol Global Trade Item Number Caution Prescription Only (applicable in the United States) Model Number CE marking Lot number 4. Description of TRIOS Serial Number 4.1 Certification and Compliance Notes Consult instructions for use CE Marking This product bears CE marking in accordance with the provisions of Medical...
  • Page 4 • Protection Tip, (S-Ptip) (S3): Protects the optical window of the scanner from damage and dirt when the scanner is not in use. When the tip The 3Shape TRIOS 3 IOS system is intended to obtain 3D digital images is detached, the scanner’s optical window (S4) is visible.
  • Page 5 • A USB license dongle (access to 3Shape software) Step 1: Connect the Ethernet cable to the pod. The cable goes into the left connection socket on the back side of the pod (the side that has two 4.7 Setting up TRIOS...
  • Page 6 4.8 Starting and Shutting Down TRIOS IOS system CAUTION When switching off the system, allow enough time for it to Please follow the steps below to start TRIOS IOS system for the first time: shut down properly before disconnecting the mains cable. Step 1: Turn the system ON.
  • Page 7: Safety And Warnings

    5. Safety and Warnings Portable RF communications equipment WARNING 5.1 Prerequisites Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) CAUTION to any part of the MOVE/MOVE+ cart system, including Read all instructions carefully, including all warnings cables specified by the manufacturer.
  • Page 8 Stress on cables System S1P (pod) only CAUTION WARNING All externally connected cables (the power cable, • Always place the scanner in its intended mount on the network cable, and the handheld scanner cable) pod when not in use. must never be subjected to pulling stress. •...
  • Page 9: Cleaning, Disinfection, And Sterilization

    USB service connection options and requirements System S1P (pod) only WARNING System S1A (MOVE+ cart) only Before each patient, make sure any PC connected to the system and its peripherals are disinfected. WARNING It is recommended to use medical-grade peripherals Only connect bus-powered USB devices without external (e.g., keyboards and mice) that can be disinfected easily.
  • Page 10 High-level disinfection with Wavicide-01 Solution as an Use an autoclave program that dries the pouched tip before opening alternative to sterilization has been validated by 3Shape. the autoclave. A class B Vacuum autoclave according to EN13060 is The Wavicide-01 Solution is not allowed and/or recognized required for sterilization of the scanner tip.
  • Page 11: Disposal

    7.2 Disposal of the TRIOS IOS system CAUTION For disposal of the TRIOS IOS system, please contact your 3Shape reseller, who Do not spray directly onto any parts of the system, can inform you about the local requirements for disposal of electronic waste.
  • Page 12: Technical Maintenance

    In case of any serious incident that has occurred in relation to the use of the CAUTION device, please report the incident to 3Shape at email: support@3shape.com All technical service and repair of the TRIOS IOS system and you can also report it to the competent authority of the state in which must be performed by 3Shape or 3Shape-authorized...
  • Page 13 Let’s change dentistry together 3Shape TRIOS A/S Holmens Kanal 7 1060 Copenhagen More information at 3Shape.com Denmark...