Installing Flow Sensors
Installing Flow Sensors
1. If you are using adapters, snap them into the sensor in the orientation shown in Figure 3.
2. Cut each original run hose to about 8 inches (20.32 cm), or other length that allows for a
downward flow and appropriate fit once the sensor and adapters (if applicable) are
installed.
3. Orient the sensor (153510-000022 or 153510-000066) so that the flow arrows point
away from the manifold. Slide a sensor assembly on each final run hose piece until the
hose piece hits the backstop inside of the sensor assembly. Make sure that you install
the sensor so that the auditory tube is pointing up.
4. Secure using a hose clamp (356060-000025) between the two ridges on the end of the
sensor assembly, ensuring that the hose clamp lays flat and the head (worm drive) does
not hit the sensor ridge, as shown in Figure 4.
5. Slide the remaining final hose piece into the lower end of the sensor assembly until the
hose piece hits the backstop inside of the sensor assembly. Secure using a hose clamp
between the two ridges on the end of the sensor assembly, ensuring that the hose clamp
lays flat and the head (worm drive) does not hit the sensor ridge.
600840-000069, rev 1.0

Figure 3: Installing flow sensors on a manifold