Condensation
C A U T I O N

Bench Space

The module dimensions and weight (see
the module on almost any desk or laboratory bench. It needs an additional
2.5 cm (1.0 inches) of space on either side and approximately 8 cm (3.1 inches)
at the left side for air circulation and electric connections.
If the bench shall carry a complete HPLC system, make sure that the bench is
designed to bear the weight of all modules.
Do not cover ventilation slots of the module.
Condensation within the module
Condensation will damage the system electronics.
Do not store, ship or use your module under conditions where
temperature fluctuations could cause condensation within the
module.
If your module was shipped in cold weather, leave it in its box
and allow it to warm slowly to room temperature to avoid
condensation.