Before service work the system must be disconnected from a current and the network.
Only touch the glass tubes with gloves.
Eyes and skin must be protected from UVC radiation. While working near ignited tubes,
please wear protective goggles, gloves, and covering clothing.

Operating time of the UVC tubes

UVC tubes have a limited life span. They are subject to a continuous aging
process. The life of a UVC tube is, with normal usage, ca 12,000 h (Output 25 %
±5 %
The effective radiation is far outside the visible range, and the UVC emissions
decreases continuously. It is recommended to add the operating conditions and
replacement date in a maintenance plan.
In order to prevent fingerprints and fat traces on the tube, we recommend to always wear cloth gloves
when working with the tubes.
orca GmbH
To the nominal UVC emissions of the respective UVC source).
Checking the O-rings when replacing tubes
While replacing tubes, check the seals, clean them, and replace
them if damaged.
www.uvpro.de
Recirculating Air Disinfection System V-lab
Seite 22 von 28