bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
packaging of the disinfectant associated with the concentration chosen for the solution.
on
- Remove the disinfected parts and rinse abundantly with warm potable water.
ry,
- Dispose of the solution following the instructions provided by the disinfectant manufac-
of
turer.
method B: Disinfect the accessories 1-2-3-4-8-10-11-13.1-13.2-13.3-13.4 by boiling
z-
them in water for 10 minutes; use demineralised or distilled water to prevent calcium de-
he
posits.
al
method C: Disinfect the accessories 1-2-3-4-8-10-11-13.1-13.2-13.3-13.4 with a hot
of
steam steriliser for baby-bottle (not the microwave type). Perform the process faithfully fol-
n.
lowing the instructions of the steriliser. To ensure that the disinfection is effective, choose
a steriliser with an operating cycle of at least 6 minutes.
After having disinfected the accessories, shake them vigorously and place them on a pa-
per towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
CLEANING IN A CLINICAL OR HOSPITAL SETTING - DISINFECTION AND STERILISATION
om
Before disinfection or sterilisation, sanitise the nebuliser cup and the accessories, choosing
one of the methods described below.
method A: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 under pota-
ny
ble hot water (approximately 40°C) with a gentle, non abrasive dish detergent.
method B: Sanitise accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 in the dish-
washer with a hot cycle.
zle
DISINFECTION
he
Accessories 1-2-3-4-5-6-7-8-10-11-13.1-13.2-13.3-13.4 can be disinfected.
The disinfectant must be an electrolytic chloroxidizer (active principle: sodium hypochlo-
rite) specific for disinfecting, which is available in any pharmacy.
Implementation:
- Fill a container big enough to hold all of the parts to disinfect with a solution of potable
of
water and disinfectant, according to the proportions indicated on the packaging of the
disinfectant.
o-
- Completely immerse each part in the solution, taking care to avoid the formation of air
bubbles on the parts. Leave the parts immersed for the amount of time indicated on the
h-
packaging of the disinfectant associated with the concentration chosen for the solution.
- Remove the disinfected parts and rinse abundantly with warm potable water.
rs-
- Dispose of the solution following the instructions provided by the disinfectant manufac-
der
turer.
If you want to also perform the STERILISATION, jump to the STERILISATION paragraph.
ON
After having disinfected the accessories, shake them vigorously and place them on a pa-
per towel. Alternatively, dry them with a jet of hot air (for example, from a hair dryer).
er
STERILISATION
Accessories 1-2-3-4-8-11-13.1-13.2-13.3-13.4 can be sterilised.
Device: Fractionated vacuum overpressure steam steriliser in accordance with EN 13060.
he
Implementation: Wrap every single part to be treated with a sterile barrier system or
packaging in accordance with Norm EN 11607. Place the packed components in the steam
.
steriliser. Run the sterilisation cycle according to the operating instructions of the device by
o-
selecting a temperature of 134 °C and a time of 10 minutes first.
Storage: Store the sterilised parts as per the instructions for use of either the sterile barrier
system or packaging.
ble
The sterilisation procedure is validated in its conformity to ISO 17665-1.
he
At the end of each use store the accessories in a dry place away from dust.
For the "CLEANING SANITISATION DISINFECTION STERILISATION" of accessories 9 fol-
air
low the instructions in the instruction manual.