Handling the AGFA CR Detectors, Plates and Cassettes
7. Wait approximately 10 minutes until the surface is dry before putting the
plate back in the cassette.
8. Put the image plate back into the cassette.
Verify that the white phosphor side is oriented to the black tube side of the
cassette and that the shutter does not scratch over the image plate.
1.
2.
2322D EN 20180409 1008
WARNING:
The use of Agfa CR Phosphor Plate Cleaner and Polynit
wipes for CR MM3.xR is restricted to image plates that have
a batch number starting with the letter "C" or higher. For CR
MM3.xR image plates with a batch code starting with the
letter "B", use only the dedicated mammography PROSAT
wipers purchased from Agfa or an Agfa certified dealer to
clean the plate.
Note: In a mixed environment with old and new CR MM3.xR
plates it is recommended to use only PROSAT wipers
purchased from Agfa or an Agfa certified dealer to clean the
plates.
WARNING:
If CR Phosphor Plate Cleaner is used, do not pour it directly
onto the image plate.
White phosphor side
Black tube side of the cassette