Warnings, Precautions And Safety Considerations - Nuheara IQbuds2 PRO Instructions For Use Manual

Hide thumbs Also See for IQbuds2 PRO:
Table of Contents

Advertisement

Warnings, Precautions and Safety Considerations

For safe and effective operation of the IQbuds 2 PRO hearing aid, please familiarize yourself with the
following information. Please read and keep all safety and use instructions.
Note: The IQbuds 2 PRO hearing aid is designed for personal and home health environment use.
Warnings
1. Use appropriate hearing protection in loud environments. Nuheara IQbuds 2 PRO hearing
aids do not function as hearing protection devices.
2. Remove the hearing aids from ears if you encounter overly loud sounds or they become
uncomfortable.
3. Check before using the hearing aids in areas where electronics or wireless devices are
restricted. This may cause interference. [
4. Do not resell the hearing aids or share them with others. They are a personal device intended
for use by a single user.
5. Do not submerge the hearing aids or Charge Case in water, wear in the shower, or wear while
participating in water sports, e.g., swimming, waterskiing, surfing, etc. Exposure to water
may damage the hearing aids.
6. Keep the product away from fire and excessive heat (e.g., from storage in direct sunlight, near
candles, or in a hot car) to avoid burns or damage to product.
7. IQbuds 2 PRO hearing aids and Charge Case should not be used in temperatures exceeding
104˚F (40˚C), placed on surfaces that exceed these temperatures, or used when conditions are
expected to exceed these temperatures. Doing so could result in overheating, fire or injury.
8. The hearing aids do not comply with requirements for explosive or oxygen-enriched
atmospheres. Do not use the hearing aids in such areas.
9. Check first before using the hearing aid system in areas where electronics or wireless devices
are restricted.
EMC Warnings
Do not use near active HF surgical equipment and the RF shielded room of an ME system for
magnetic resonance imaging, where the intensity of EM disturbances is high.
Do not change any cable without authorization of the manufacturer. Use of cables other than
those specified or provided by the manufacturer of this appliance could result in increased
electromagnetic emissions or decreased electromagnetic immunity of this appliance and result
in improper operation.
Use of this appliance adjacent to or stacked with other device should be avoided because it
could result in improper operation. If such use is necessary, this appliance and the other
device should be observed to verify that they are operating normally.
Portable RF communications equipment (including peripherals such as antenna cables and
external antennas) should be used no closer than 30 cm (12 inches) to any part of the IQbuds
2 PRO hearing aid, including cables specified by the manufacturer. Otherwise, degradation of
this equipment could result.
For maintaining the basic safety and essential performance with regard to electromagnetic
disturbances for the expected service life, please take precautions if the use location is near
(e.g. less than 1,5 km from) from AM, FM or TV broadcast antennas.
8