Apera Instruments ZenTest PH60F-Z Instruction Manual

Smart flat ph tester
Hide thumbs Also See for ZenTest PH60F-Z:

Advertisement

Quick Links

ZenTest™ PH60F-Z Smart Flat pH Tester
(pH/Temp./ORP)
Instruction Manual
APERA INSTRUMENTS (Europe) GmbH
www.aperainst.de
- 1 -

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ZenTest PH60F-Z and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Apera Instruments ZenTest PH60F-Z

  • Page 1 ZenTest™ PH60F-Z Smart Flat pH Tester (pH/Temp./ORP) Instruction Manual APERA INSTRUMENTS (Europe) GmbH www.aperainst.de - 1 -...
  • Page 2: Table Of Contents

    Contents 1. Introduction ........................- 3 - 2. What’s in the Kit ......................- 4 - 3. Battery Installation ......................- 4 - 4. Keypad Functions ......................- 5 - 5. Preparation before Use ....................- 5 - 6. pH Calibration ........................- 5 - 7.
  • Page 3: Introduction

    1. Introduction Thank you for choosing Apera Instruments PH60F-Z Smart Flat pH Tester. Please carefully read this manual before using the product. Search “zentest smart” in Apple App Store or Google Play App Store to download the latest App for your tester. Turn on the Bluetooth of your phone, go to ZenTest™ App, tap on the upper right corner, then select your tester to connect.
  • Page 4: What's In The Kit

    2. What’s in the Kit PH60F-Z Tester Calibration Vials 3. Battery Installation Please install batteries according to the following steps. *Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) - 4 -...
  • Page 5: Keypad Functions

    4. Keypad Functions ■ Short press------ < 2 seconds ,Long press------- > 2 seconds 1. When turned off, short press to turn on the tester; long Battery Cap press to enter parameter setting. 2. In calibration mode or parameter setting, short press to return to measurement mode.
  • Page 6 Pour certain amount of pH 7.00 and pH 4.00 buffer solution into the corresponding calibration bottles (to about half volume of the bottle); Long press to enter calibration mode; Short press to exit. Insert the probe into pH 7.00 buffer solution, make a quick stir, and hold still.
  • Page 7: Ph Measurement

    For the self-diagnosis information, please refer to the table below: Symbol Self-Diagnosis information Potential problems and how to fix 1. Make sure the probe is fully immersed in the calibration solution. 2. Check if calibration solution is expired or polluted. 3.
  • Page 8: Orp Measurement

    is stabilized ( appears and stays on screen). PH60F Flat probe testers are mostly for flat surface sample test. 1) For skin test: skin should be without sweat or dirt, nor be overly cleaned (do not use facewash products before testing) to avoid affecting measurement results, dampen skin with some distilled water, slightly force flat probe onto the skin, get readings after value stabilized.
  • Page 9: Parameter Setting

    affected at all. 4) NEVER store the probe in pure water like tap, RO, distilled, or deionized water as they could damage the pH probe. If this happens, immediately soak the pH probe Apera 3M KCL soaking solution overnight, then re-calibrate it before using. Pure water is only for rinsing the probe. 11.
  • Page 10: Technical Specifications

    appear in the lower right corner of the LCD screen in 3 days to remind you to perform calibration, also in the ZenTest App there will be a pop-up reminder. After calibration is finished or the reminder setting is cancelled in the ZenTest App, the Er6 icon will disappear. 6) pH Back to Factory Default (P08) –...
  • Page 11: Probe Replacement

    14. Probe Replacement pH probes don’t last forever. Every probe will eventually age and fail even if you don't use it that often. The typical service life of Apera probes is 12-24 months depending on the frequency of usage and how well you keep it clean and properly stored. We recommend replacing your probe every 1 to 2 years to ensure the best performance To replace a probe: 1) Take off the probe cap;...
  • Page 12: Warranty

    We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS (Europe) GmbH, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS (Europe) GmbH for a period of TWO YEARS (SIX MONTHS for the probe) from the delivery.