Endress+Hauser Liquistation CSF39 Brief Operating Instructions
Endress+Hauser Liquistation CSF39 Brief Operating Instructions

Endress+Hauser Liquistation CSF39 Brief Operating Instructions

Automatic sampler for liquid media
Hide thumbs Also See for Liquistation CSF39:

Advertisement

Quick Links

KA01184C/07/EN/02.16
71314891
Valid as of version
01.06.00
Products
Brief Operating Instructions
Liquistation CSF39
Automatic sampler for liquid media
These Instructions are Brief Operating Instructions; they are
not a substitute for the Operating Instructions pertaining to
the device.
Detailed information on the device can be found in the
Operating Instructions and in the other documentation
available at:
• www.endress.com/device-viewer
• Smart phone/tablet: Endress+Hauser Operations App
Solutions
Services

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Liquistation CSF39 and is the answer not in the manual?

Questions and answers

Summary of Contents for Endress+Hauser Liquistation CSF39

  • Page 1 These Instructions are Brief Operating Instructions; they are not a substitute for the Operating Instructions pertaining to the device. Detailed information on the device can be found in the Operating Instructions and in the other documentation available at: • www.endress.com/device-viewer • Smart phone/tablet: Endress+Hauser Operations App...
  • Page 2 Liquistation CSF39 Order code 00X00-XXXX0XX0XXX Ser. No.: X000X000000 TAG No.: XXX000 Serial number www.endress.com/deviceviewer Endress+Hauser Operations App A0023555 Endress+Hauser...
  • Page 3: Table Of Contents

    Liquistation CSF39 Table of contents Table of contents Document information ............4 Warnings .
  • Page 4: Document Information

    Document information Liquistation CSF39 Document information Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious injury. Causes (/consequences) Consequences of non-compliance (if applicable) ‣...
  • Page 5: Documentation

    Liquistation CSF39 Document information Documentation The following manuals which are available on the product pages on the internet complement these Operating Instructions: • Operating Instructions for – Device description – Commissioning – Operation – Software description (excluding sensor menus; these are described in a separate manual - see below) –...
  • Page 6: Basic Safety Instructions

    Designated use Liquistation CSF39 is a stationary sampler for liquid media. The samples are taken discontinuously using a vacuum pump or peristaltic pump and are then distributed into sampling containers and refrigerated.
  • Page 7: Operational Safety

    Liquistation CSF39 Basic safety instructions Operational safety Before commissioning the entire measuring point, verify that all connections are correct. Ensure that electrical cables and hose connections are undamaged. Do not operate damaged products, and safeguard them to ensure that they are not operated inadvertently.
  • Page 8: Incoming Acceptance And Product Identification

    Incoming acceptance and product identification Liquistation CSF39 Incoming acceptance and product identification Incoming acceptance Verify that the packaging is undamaged.  Notify your supplier of any damage to the packaging. Keep the damaged packaging until the matter has been settled.
  • Page 9: Scope Of Delivery

    Liquistation CSF39 Incoming acceptance and product identification Scope of delivery The scope of delivery comprises: • 1 Liquistation CSF39 with: – The ordered bottle configuration – Optional hardware • Accessories kit For peristaltic or vacuum pump: Connection nipple for suction line with various angles (straight, 90°), Allen key (for version with vacuum pump only) •...
  • Page 10: Installation

    Installation Liquistation CSF39 Installation Installation conditions 4.1.1 Dimensions 625 (24.6) 471 (18.5) 753 (29.7) 816 (32.1) 355 (14.0) 125 (4.9) A0024494  1 Dimensions of Liquistation CSF39 plastic version with stand, dimensions in mm (in) Suction line connection Endress+Hauser...
  • Page 11 Liquistation CSF39 Installation 624 (24.6) 467 (18.4) 743 (29.3) 816 (32.1) 355 (14.0) 125 (4.9) A0024495  2 Dimensions of Liquistation CSF39 stainless steel version with stand, dimensions in mm (in) Suction line connection Endress+Hauser...
  • Page 12 Installation Liquistation CSF39 4.1.2 Installation site For version with pump A0024411  3 Installation conditions for the Liquistation Correct  The suction line must be routed with a downward slope to the sampling point. Incorrect  The sampler should never be mounted in a place where it is exposed to aggressive gases.
  • Page 13 Liquistation CSF39 Installation 4.1.3 Mechanical connection Foundation plan 730 (28.7) 311 (12.2) 305 (12.0) 255 (10.0) 302 (11.9) 214 (8.43) 786 (30.9) A0024406  4 Foundation plan Fasteners (4 x M10) Cable inlet Outlet for condensate and overflow > DN 50 Sample supply from below >...
  • Page 14 Installation Liquistation CSF39 4.1.4 Connection for suctioning samples • Maximum suction height: – Vacuum pump: Standard 6 m (20 ft) Optional 8 m (26 ft) – Peristaltic pump: standard 8 m (26 ft) • Maximum hose length: 30 m (98 ft) •...
  • Page 15 Liquistation CSF39 Installation Note the following when erecting the device: • Always route the suction line so that it slopes upwards from the sampling point to the sampler. • The sampler must be located above the sampling point. • Avoid siphoning effects in the suction line.
  • Page 16: Installation

    Installation Liquistation CSF39 Installation 4.2.1 Connecting the suction line at the side on version with pump When installing the device, take the installation conditions into account. Route the suction line from the sampling point to the device. Screw the suction line onto the device' s hose connection.
  • Page 17 Liquistation CSF39 Installation Connecting the suction line on version with vacuum pump A0013708 A0013707  7 Suction line connected from below  6 Connecting the suction line from the side (as- delivered state) Hose Fixing clip for hose gland Thread adapter nut Hose gland Unscrew the thread adapter nut (item 3).
  • Page 18: Post-Installation Check

    Installation Liquistation CSF39 Connecting the suction line on version with peristaltic pump A0013705 A0013706  8 Connecting the suction line from the side (as-  9 Suction line delivered state) Small thread adapter nut Hose Thread adapter nut Hose gland Unscrew the thread adapter nut (item 3) and the hose fitting (item 4) from the side panel.
  • Page 19: Electrical Connection

    Liquistation CSF39 Electrical connection Electrical connection Connecting the sampler WARNING Device is live Incorrect connection may result in injury or death ‣ The electrical connection may be performed only by an electrical technician. ‣ The electrical technician must have read and understood these Operating Instructions and must follow the instructions contained therein.
  • Page 20 Electrical connection Liquistation CSF39 5.1.3 Removing the rear panel of the dosing compartment Open the door of the dosing compartment. Using an 5mm Allen key, release the rear panel by turning the lock clockwise.  CLOSE OPEN A0012803 Lift up the rear upper panel and pull it off towards the back.
  • Page 21 Liquistation CSF39 Electrical connection 5.1.4 Removing the rear panel of the sampling compartment A0012825 ‣ Remove the bolt on the rear of the dosing compartment.  A0012824 ‣ Remove the bolt on the rear panel. 5.1.5 Removing the cover on the power unit...
  • Page 22 Electrical connection Liquistation CSF39 A0012831 Release the screw with an Allen key (5 mm). Remove the cover of the power unit from the front. When reassembling make sure that the seals are seated correctly. 5.1.6 Power supply terminal assignment The power supply is connected via plug-in terminals.
  • Page 23 Liquistation CSF39 Electrical connection A0013237  11 Terminal assignment Assignment: 100 to 120 V/200 to 240 V AC ±10 % Assignment: 24 V DC +15/-9 % Rechargeable batteries (optional) Internal 24 V voltage Fuses (only for batteries) Endress+Hauser...
  • Page 24 Electrical connection Liquistation CSF39 5.1.7 Cable terminals Plug-in terminals for Memosens and PROFIBUS/RS485 connections Insert the cable until the limit stop Press the screwdriver against the clip Remove the screwdriver (closes the (opens the terminal) terminal) After connection, make sure that every cable end is securely in place. Terminated cable ends, in particular, tend to come loose easily if they have not been correctly inserted as far as the limit stop.
  • Page 25: Connecting Modules And Sensors

    Liquistation CSF39 Electrical connection Connecting modules and sensors 5.2.1 Connection compartment in the controller housing A0013238 A0012843 1 L base module Sampler controller The controller housing has a separate connection Display cover open, version with L base module compartment. Release the six cover screws to open the connection compartment: ‣...
  • Page 26 Electrical connection Liquistation CSF39 5.2.2 E base module 24 VDC Power A0016535  12 E basic module Indicator LEDs Slot for display cable Voltage connection Service interface Alarm relay connection Connections for 2 Memosens sensors (optional) Power supply for digital fixed cable sensors...
  • Page 27 Liquistation CSF39 Electrical connection 85 86 24VDC Alarm Sensor supply Power – (internal) – – 0/4 ... 20 mA Display Service HART – Sensor A0016537  13 E basic module wiring diagram 5.2.3 Connecting the sensors Sensor types with Memosens protocol...
  • Page 28 Electrical connection Liquistation CSF39 Sensor Sensor A0023038 A0023039  14 sensors without additional supply voltage  15 sensors with additional supply voltage Sensor Sensor 1 Sensor 2 Sensor A0016197  16 sensors with and without additional supply voltage on 2DS sensor module...
  • Page 29 Liquistation CSF39 Electrical connection Sensor connection ‣ Guide the sensor cable via the rear panel to the controller housing towards the front. →  20 and →  21  A0016360  17 Gland to the controller Only use terminated genuine cables where possible.
  • Page 30 Electrical connection Liquistation CSF39 Connecting the ferrules of the sensor cable to the E base module ‣ Ground the outer shield of the cable via the metal gland to the left of the E base module.  32 31 A0016356 ...
  • Page 31 Liquistation CSF39 Electrical connection Analog inputs and binary inputs/outputs 125 123 124 225 223 224 191 192 291292 145 146 245 246 A0012988  20 Position of the terminals Analog inputs 1 and 2 Binary inputs/outputs Analog inputs A0012989  21...
  • Page 32 Electrical connection Liquistation CSF39 CSXxx CMxx A0028652  22 With two-wire transmitter, e.g. Liquiline M CM42 CSXxx FMU90 A0028653  23 With four-wire transmitter, e.g. Prosonic S FMU90 Binary inputs A0013381  24 Assignment of binary inputs 1 and 2...
  • Page 33 Liquistation CSF39 Electrical connection When connecting to an internal voltage source, use the terminal connection on the rear of the dosing compartment. The connection is located on the lower terminal strip (on the far left, + and -), (→  22)
  • Page 34 Electrical connection Liquistation CSF39 Binary outputs with relay option A0016343  28 Relay Binary output 1 Binary output 2 The left relay is activated with binary output 1, while the right relay is activated by binary output 2. A0016348  29...
  • Page 35: Terminal Assignment For Input/Output Signals

    Liquistation CSF39 Electrical connection Terminal assignment for input/output signals Input signals • 2 analog signals 0/4 to 20 mA • 2 binary signals > 100 ms pulse width or edge Signals of digital sensors with Memosens protocol (optional) Output signals 2 binary signals >...
  • Page 36: Connecting Additional Inputs, Outputs Or Relays

    Electrical connection Liquistation CSF39 Connecting additional inputs, outputs or relays WARNING Module not covered No shock protection. Danger of electric shock! ‣ If you are modifying or extending your hardware, always fill the slots from top to bottom. Do not leave any gaps.
  • Page 37 Liquistation CSF39 Electrical connection 5.4.2 Current outputs – – – – – –  32 Module  33 Wiring  34 Module  35 Wiring diagram diagram A maximum of 6 current outputs are supported. Endress+Hauser...
  • Page 38: Connecting Digital Communication

    Electrical connection Liquistation CSF39 Connecting digital communication 5.5.1 Module 485 128/SW Service DGND  37 Wiring  36 Module diagram LEDs on front of module Description Color Description RJ45 LNK/ACT • Off = Connection is not active • On = Connection is active •...
  • Page 39 Liquistation CSF39 Electrical connection DIP switches on front of module Factory setting Assignment 1-128 Bus address (→ "Commissioning/communication") Write protection: "ON" = configuration not possible via the bus, only via local operation Service If the switch is set to "ON" , the user settings for Ethernet addressing are saved and connection settings programmed into the device at the factory are activated: IP address=192.168.1.212, Subnet mask=255.255.255.0, Gateway=0.0.0.0, DHCP=Off.
  • Page 40 Electrical connection Liquistation CSF39 DIP switches on front of module Factory setting Assignment 1-128 Bus address (→ "Commissioning/communication") Write protection: "ON" = configuration not possible via the bus, only via local operation Service If the switch is set to "ON" , the user settings for Ethernet addressing are saved and connection settings programmed into the device at the factory are activated: IP address=192.168.1.212, Subnet mask=255.255.255.0, Gateway=0.0.0.0, DHCP=Off.
  • Page 41: Hardware Settings

    Liquistation CSF39 Electrical connection 5.5.3 Bus termination There are two ways to terminate the bus: 1. Internal terminating resistor (via DIP switch on the module board) "OFF" "ON"  40 DIP switches for internal terminating resistor ‣ Using a suitable tool, such as a tweezers, set all 4 DIP switches to the "ON" position.
  • Page 42: Ensuring The Degree Of Protection

    Electrical connection Liquistation CSF39 Set the desired bus address via the DIP switches of module 485. For PROFIBUS DP, valid bus addresses are anything between 1 and 126, and anything between 1 and 247 for Modbus. If you configure an invalid address, software addressing is automatically enabled via the local configuration or via the fieldbus.
  • Page 43: Post-Connection Check

    Liquistation CSF39 Electrical connection Post-connection check WARNING Connection errors The safety of people and of the measuring point is under threat. The manufacturer does not accept any responsibility for errors that result from failure to comply with the instructions in this manual.
  • Page 44: Operation Options

    Operation options Liquistation CSF39 Operation options Overview 6.1.1 Display and operating elements Display (with red display background in alarm condition) Navigator (jog/shuttle and press/hold EH_CSF 09:11:05 31.03.2015 function) Soft keys (function depends on menu) MODE A0025501  45 Overview of operation 6.1.2...
  • Page 45: Access To The Operating Menu Via The Local Display

    Liquistation CSF39 Operation options Access to the operating menu via the local display 6.2.1 Operating concept MODE MODE Turning the navigator: moving the cursor in the menu Pressing the soft key: selecting the menu directly Menu/Language MODE Turning the navigator: selecting a value (e.g. from a list)
  • Page 46: Configuration Options

    Operation options Liquistation CSF39 6.2.2 Locking or unlocking operating keys Locking operating keys ‣ Press the navigator for longer than 2 s.  A context menu for locking the operating keys is displayed. You have the choice of locking the keys with or without password protection. "With password"...
  • Page 47 Liquistation CSF39 Operation options 6.3.3 Numerical values • You are changing a variable. • The maximum and minimum values for this variable are shown on the display. • Set a value within this range. • Example: Menu/Operation/Display/Contrast Menu/...ration/Display/Contrast 6.3.4 Actions •...
  • Page 48 Operation options Liquistation CSF39 6.3.5 Free text • You are assigning an individual designation. • Enter a text. You can use the characters in the editor for this purpose (upper-case and lower-case letters, numbers and special characters). • Using the soft keys, you can: –...
  • Page 49 Liquistation CSF39 Operation options 6.3.6 Tables • Tables are needed to map mathematical functions or to enter irregular interval samples. • You edit a table by navigating through rows and columns with the navigator and changing the values of the cells.
  • Page 50: Commissioning

    Commissioning Liquistation CSF39 Commissioning Function check WARNING Incorrect connection, incorrect supply voltage Safety risks for staff and device malfunctions ‣ Check that all connections have been established correctly in accordance with the wiring diagram. ‣ Ensure that the supply voltage matches the voltage indicated on the nameplate.
  • Page 51: Basic Setup

    Liquistation CSF39 Commissioning Basic setup Making basic settings Device tag: Give your device any name of your choice (max. 32 characters). Set date: Correct the set date if necessary. Set time: Correct the set time if necessary. Number of bottles: Correct the set number of bottles if necessary.
  • Page 52: Sampling Programs

    Commissioning Liquistation CSF39 Sampling programs 7.4.1 Difference between program types The following box provides an overview of the differences between the Basic, Standard and Advanced program types. Basic (1 sampling program) Start condition: • Immediate activation Stop condition: • Immediate •...
  • Page 53 Liquistation CSF39 Commissioning After manual sampling, a running program can be displayed and continued with the ESC ESC  Menu/Manual sampling 15000 ml Distribution position Bottle 1 MODE A0024602-EN The sample volume for "Manual sampling" is not taken into account in the calculated bottle volumes.
  • Page 54 Commissioning Liquistation CSF39 Stop condition: after program end or continuous operation. Pressing the SAVE saves the program and ends data entry.  Example: 15000 ml 100 ml MODE A0029242-EN The program can be started. Endress+Hauser...
  • Page 56 *71314891* 71314891 www.addresses.endress.com...