Download Print this page

Radiant Innovation TH809 Quick Start Manual page 2

Ear-type thermometer

Advertisement

 This thermometer
is an adjusted mode thermometer that
(according to the result of the clinical evaluation to get the offset value)
 Enclosure Rating: IP22
 Dimensions: 141.4 x 40.4 x 43.5 mm
 Weight: 67.3 grams including battery
 Battery life: around 3,000 continuous readings.
 Expected Service Life: 4 years
The device should not submerge into any liquids and expose it to direct moisture.
There is no gender and age limitation for using infrared thermometer.
Troubleshooting:
Error Message
Problem
Device stabilization is in process.
Battery is low and no more measurements are
possible.
Measurement before device stabilization.
The ambient temperature is not within the normal
temperature range between 10
o
~104
o
F).
Error 5~9, the system is not functioning properly.
Temperature taken is higher than 42.2
Temperature taken is lower than 34
Device can not be powered on to the ready stage.
Warranty: 12 months
Manufacture Date: as the serial number (please open the battery cover, it is shown on the inside of the device.)
Ex.SN:E512A000001, the first "E" is External, the second number "5" is the last number of manufacture year, the third and the
fourth number "12" is the manufacture month, the others is the serial number.
Note: The thermometer is calibrated at the time of manufacture. If you question calibration mode, the accuracy of
temperature measurements or unexpected events at any time, please contact the dealers or nearest service
address.
Please report to the manufacturer and the competent authority of the Member State in which you are established
about any serious incident that has occurred in relation to this device.
Warning: No modification of this equipment is allowed.
Please read the instructions for use
IMPORTER
Company Name?
Company Name?
General Information: Address? Website?
Radiant Innovation Inc.
Tel: +886 (3) 5644185 Fax: +886 (3) 5644170
E-mail: MKT@radiantek.com.tw Http://www.radiantek.com.tw
Add: 1F, No.3, Industrial East 9
Road, Science-Based Industrial Park,
th
HsinChu, Taiwan 300.
converts the ear temperature to display its "oral equivalent."
This is not an AP or APG product.
Solution
Wait until
stops flashing.
Replace the battery.
Wait for "Er1" to disappear.
Allow the thermometer to rest in a room at least 15
C and 40
o
C (50
o
F
minutes at room temperature: 10
~104
o
F).
Unload the battery, wait for 1 minute and repower it.
If the message reappears, contact the retailer for
service.
Check the integrity of the probe cover and take a new
o
C (108.0
o
F).
temperature measurement.
Make sure the probe cover is clean and take a new
o
C (93.2
o
F).
temperature measurement.
Change with a new battery.
BF type applied part
Date of issue: 2020-06-15
Ref No.: 062020
Medical Technology Promedt Consulting Gmb
Add: Altenhofstrasse 80, D-66386 St. Ingbert, Germany
o
C and 40
o
C (50
o
F

Advertisement

loading