• Pour sealing compound down in between the two fiber rope dams filling the remaining space. Pour slowly,
being careful not to trap air bubbles. Immediately wipe off any spilled sealing compound.
• Screw conduit domed-cover back onto conduit opening.
NOTE: INITIAL SETUP OF SEALING COMPOUND WILL OCCUR IN APPROXIMATELY 30 MINUTES
HOWEVER, THE SEALING COMPOUND REQUIRES A MINIMUM OF 8 HOURS ABOVE 32ºF TO
DEVELOP SUFFICIENT STRENGTH TO WITHSTAND EXPLOSIONS.
Note: Wires must be kept separated as shown in the diagram before addition of the sealing compound.
14
I
NSTALLATION