Omega HHF143 User Manual page 13

Handheld rotating vane thermo-anemometer
Hide thumbs Also See for HHF143:
Table of Contents

Advertisement

Note:
When using analog outputs, there is an
analog output voltage. This is in addition to the normal measurement error.
► For example, an air velocity reading of 500 FPM would normally have an
accuracy of ±1 % of reading ±1 digit (±6 FPM) when the data is viewed
on the LCD display.
► With the additional error associated with the analog output voltage, the
effective accuracy of the analog output for this air velocity measurement
will be ±2% ofreading ±1 digit (±11 FPM).
Custom Analog Voltage Outputs are also available - contact Omega for details.
Analog Output circuit block diagram and connector pin assignment
additional ±1
error
%
in the
ANALOG OUTPUT CONNECTOR
PINOUT DIAGRAM
Mating Cable Connector:
Binder Part No. 99-0413-00-0�'
Rev 1.8, 1 0-Jun-2008
Page 12 of24